WIGGENS WH220-HT
e.g. T2 Set to 151
e.g. The corresponding AD value of S2
is 1468
thermometer)
Wait for the temperature to reach and equilibrium
refer to the shows value on third-party thermometer and press the Up
Button and Down Button to set the value of S2
now the display shows its corresponding AD read value of S2(e.g. The
corresponding AD value is 1468
Press and hold the START / STOP Button and observe the value change
on the display
Release the START / STOP button when the temperature fluctuation is
less or equal to ±1
Press the SELECT Button on the operation panel to complete the
calibration process.
The display will show the model and software version again.
Caution!
The T1 and T2 temperatures can be between 20°C to 200°C, since the maximum
measured temperature of the PT100 temperature sensor can be up to 200°C.
The temperature environment between T1 and T2 should be greater than 50°C to
ensure the accurately calibration result.