![Wiggens WH220-HT Скачать руководство пользователя страница 14](http://html1.mh-extra.com/html/wiggens/wh220-ht/wh220-ht_operating-manual_979998014.webp)
Operating Manual
4.2.2. The dimensions of the digital hot plates/stirrer
(
after connect PT100Teperature sensor
)
CAUTION!
Do not use voltages that are higher or lower than 10% of the voltage specified on the
label, which is on the backside of the instrument.
Keep the power cord and temperature sensor cable off of the hot plate while heating.
Put the solution on the top plate before operating the instrument.
Heating corrosive liquids under poor ventilation hoods will shorten the life of the
electronic components inside the instrument.
Upon the first heating operation, a particular smell and white smoke can appear. This is
normal. Put the instrument under a fume hood and moderately heat for about one hour
until the smell and smoke fully disappear.
If toxic gases are released, air circulation must be kept.
The safe temperature limit must always be set to at least 25
℃
lower than the fire point
of the media used.
.
48.5c
m
24cm
24.5c
m
14