Instructions for Operation and Care of Thermo-Therapy Dry Heat Therapy
7
Models TT-101, and TT-101L are furnished with a hospital
grade plug and require up to 13 amps for proper operation.
Therefore, as a minimum, a 15-amp breaker and GFI
protected outlet should be made available for electric
power.
Model TT-202 is also furnished with a hospital grade plug
but can require up to 17 amps of electric power. Therefore,
20 amp service and a GFI protected outlet should be
provided.
Since these units produce large volumes of exhausted
air, some minimal dusting may occur. Therefore, good
ventilation must be considered. Patients and operators
sensitive to small amounts of discharged Celstim dust
should be provided with a mask and/or the treatment ports
and the extremity being treated could be covered with a
damp towel.
Celstim on any surface, particularly on tile floor is extremely
slippery and caution must be carefully exercised. Again,
particularly tile floors must be kept free of this material – a
frequently swept, thick carpet under the unit is ideal. Also,
the floor near the air intake ports must be kept clean to
prevent small dirt, dust and Celstim particles from entering
the motor chamber.
Also, since all the units’ blowers, particularly on the TT-101
and TT-101L, and even though they are sound baffled to the
best of our ability, can be considered noisy. Consideration
should be made for treatment in a separate room or at the
end of a large room.
RECEIVING INSTRUCTIONS
All Thermo-Therapy models are shipped complete with the
proper amount of sterilized Celstim and as soon as they are
carefully unpacked they are “ready to go”.
If shipping damage has occurred, notify Whitehall
Manufacturing immediately so a proper claim can be made
and replacement or repair secured.
SPECIFICATIONS
TT-101 – the outside dimensions are 19-1/2” (50 cm) L
x 16-1/2” (41 cm) H x 10-1/2” (27 cm) When placed on
the optional stand, the bottom is 13-1/2” (34 cm) and the
top is 28” (71 cm) from the floor. The Celstim capacity is
approximately 10 lbs. (4.5 kg) and the treatment chamber
is approximately 7-1/2” (19 cm) H x 9” (23 cm) W x
15-1/2” (39 cm) L. The blower is driven by a variable speed
controlled motor. Variable heat is provided by a finned, dual
safety switched 900 watt heater.
TT-101L – the outside dimensions are 23” (58 cm) L x 18”
(46 cm) H x 10-1/2” (27 cm) W. When placed on optional
stand, the bottom is 13-1/2” (34 cm) and the top is 31”
(79 cm) from the floor. The Celstim capacity is 12 lbs.
(5.4 kg) and the treatment chamber is 8-1/2” (22 cm) H x
9” (23 cm) W x 18-1/2” (47 cm) L. The blower motor and
heater are the same for the TT-101.
TT-202 – the outside dimensions are 29” (74 cm) L x 35”
(89 cm) H x 16-1/2” (42 cm) W. The locking casters are at
the extreme width of the unit and are mounted on thick steel
plates to protect the units’ bottom from damage when there
are uneven floors. The Celstim capacity is approximately
40 lbs. (18 kg) and the treatment compartment is 22”
(56 cm) L and 15-1/2” (39 cm) W x 13” (33 cm) H.
There are two blower motors combined with the 900 watt
heater on model TT-202. Interestingly, the blowers on the
TT-202 are individually controlled and through independent
speed manipulation can provide varying air flow patterns in
the treatment chamber.
ELECTRICAL REQUIREMENT
TT-101 and TT-101L – 120V/15 Amp/60 Hz.
TT-202 – 120V/20 Amp/60 Hz.
REQUIREMENTS FOR INSTALLATION