Whitehall Para-Therapy PT-18-S Скачать руководство пользователя страница 5

Step 1–

Remove your new 

Para-Therapy 

unit from its packaging and place unit on 

a suitable level surface like a sturdy tabletop or countertop. Remove 
stainless steel cover and remove any accessories and literature from 
inside of tank.

NOTE: DO NOT PLUG UNIT IN UNTIL STEP 2 IS COMPLETED!

Step 2–

Add paraffin to unit. For Model PT-6-S add six (6) one pound bags of 
genuine 

Para-Therapy 

paraffin chips and for Model PT-18-S add eighteen 

(18) one pound bags of paraffin chips. Chips sometimes melt together 
during shipment but can be easily separated when putting in unit. (Note: If 
you are using paraffin by others and it is not in chip form, but one pound 
bars, make sure each bar is placed individually into unit so they come in 
contact with the bottom of the tank so they will melt properly.)

Step 3–

Your 

Para-Therapy 

unit is now ready to plug in. When unit is plugged in, the 

indicator light on the front of unit will come on indicating unit is under power. The 
six pound unit will take 4 to 5 hours and the eighteen pound unit will take 10 to 
12 hours to melt all the paraffin completely. Once paraffin is completely melted 
your unit is ready to use.

Operating Tips

When do I need to add Paraffin?

Add paraffin when the paraffin level in your tank drops below a depth of 3”. Using 
unit without sufficient paraffin in the tank can cause major temperature fluctuations 
in the paraffin bath.

Why is there foreign material on the bottom of the tank?

The material on the bottom of the tank is contamination such as dirt or dead skin 
which builds up after prolonged use or by placing used paraffin back in the unit after 
treatment. When this happens, we recommend removing used paraffin, cleaning 
unit and adding new paraffin. 

Why is there a film of solidified paraffin on the top surface of my tank?

The paraffin has lost some of its oil through normal usage. Add paraffin oil (food 
grade mineral oil) as required or replace paraffin.

REMARQUE: NE PAS BRANCHER L'APPAREIL AVANT D'AVOIR 
TERMINÉ L'ÉTAPE 2 !

3

WHITEHALL MANUFACTURING

 • P.O. BOX 3527 • City of Industry, CA 91744-0527 USA

Phone (800) 782-7706 • (626) 968-6681 • Fax (626) 855-4862 • Web: www.whitehallmfg.com

Operating Instructions

Instructions for Operation and Care of Paraffin Baths

Содержание Para-Therapy PT-18-S

Страница 1: ...ra Therapy Revised April 2017 5900 100 000 INSTRUCTIONS FOR OPERATION AND CARE OF 800 782 7706 626 968 6681 www whitehallmfg com P O Box 3527 City of Industry CA 91744 0527 U S A Division of Acorn Eng...

Страница 2: ...eservoir ou de les appuyercontre celui ci l RETIRER tout emballage accessoire et feuillet se trouvant l int rieur de votre nouvel appareil avant de l utiliser EFFECTUER LA MISE LA TERRE de votre appar...

Страница 3: ...ams and Specifications Model PT 6 S Page 6 Model PT 18 S Page 8 Repair Parts Model PT 6 S Page 6 Model PT 18 S Page 8 Exploded View Model PT 6 S Page 7 Model PT 18 S Page 9 Troubleshooting Guide Page...

Страница 4: ...ns 13 L x 7 W x 7 3 4 D 330 L x 180 W x 195 D Inner Tank Dimensions 11 3 4 L x 5 3 4 W x 5 3 4 D 300 L x 145 W x 145 D Tank Capacity 6 lbs 2 72 kgs Electrical Requirements Inlet Voltage 120 VAC Freque...

Страница 5: ...een pound unit will take 10 to 12 hours to melt all the paraffin completely Once paraffin is completely melted your unit is ready to use Operating Tips When do I need to add Paraffin Add paraffin when...

Страница 6: ...s of the body which cannot be dipped into the tank a soft brush may be used to apply multiple layers of paraffin to area as required for treatment 5 For maximum benefit after the paraffin has been app...

Страница 7: ...towel Carefully lift unit up an inch or two and then bump it down on the surface to break free the solid paraffin block Block should drop on to lid Lift away unit leaving the paraffin block on top of...

Страница 8: ...113 001 Euro Outer Housing 13 6505 540 000 Indicator Light 250 VAC 14 5902 113 199 Bottom Cover 15 6504 604 000 Snap In Bumper Foot 16 0331 014 000 0 203 ID x 0 734 OD Stainless Steel Washer 17 0116...

Страница 9: ...855 4862 Web www whitehallmfg com Instructions for Operation and Care of Paraffin Baths FUSE 1 FUSE 2 GND INLET L N THERMO INDICATOR HEATING ELEMENTS LIGHT STAT THERMAL CUTOUT POWER Electrical Specifi...

Страница 10: ...13 199 Bottom Cover 14 6505 540 000 Amber Indicator Light 125 V 15 0116 010 000 10 32 x 1 2 Phillips Truss Head Screw 16 0302 004 000 10 32 Hex Nut 17 6502 560 000 10 Stainless Steel Lockwasher 6504 6...

Страница 11: ...68 6681 Fax 626 855 4862 Web www whitehallmfg com Instructions for Operation and Care of Paraffin Baths 1 2 3 6 7 23 24 5 3 4 2 9 8 3 10 11 12 13 14 15 16 17 18 19 20 12 25 FUSE 1 FUSE 2 GND INLET L N...

Страница 12: ...MAL CUTOUT BURNT OR FAULTY REPLACE THERMOSTAT THERMOSTAT IS BROKEN HEATING ELEMENT IS REPLACE HEATING ELEMENTS BURNT OR FAULTY PARAFFIN WAX IS MELTING INDICATOR LIGHT REPLACE INDICATOR LIGHT INDICATOR...

Страница 13: ...stitute the sole remedy of the Buyer and the sole remedy of Whitehall under this warranty Whitehall shall not be liable under any circumstances for incidental consequential or direct charges caused by...

Отзывы: