Whitehall Para-Therapy PT-18-S Скачать руководство пользователя страница 4

Stationary Paraffin Tanks provides a simple, effective method of applying heat to the surface of body 
extremities where conditions indicate a need for increased circulation. The tank shall be for an arm or 
hand paraffin bath. 

Indications;

Paraffin bath, when prescribed, is a useful method of relieving pain caused by medically diagnosed 
arthritis, bursitis and chronic joint pain. Relaxes muscles, relieves muscle stiffness, and muscle 
spasms. Most commonly used prior to therapeutic exercise and or massage. Consult your physician if 
you have any questions. 

Contraindications;

Should not be used in the presence of open wounds, inflammatory skin conditions, growths, peripheral 
vascular disease where circulation is impaired, acute inflamation, or when there is reduced or absent of 
sensation of the extremities (such as in some cases of diabetes). Should not be used on areas subject 
to hemorrhaging or in cases involving abdominal sensitivity to heat. Discontinue use if dermatitis occurs 
due to sensitivity to paraffin wax or if wax feels too hot or cold, could indicate health problems with user. 
Consult with a physician or physical therapist before using.

Model PT-6-S Specification:

Overall Dimensions: 13" L x 7" W x 7-3/4" D (330 L x 180 W x 195 D) 
Inner Tank Dimensions: 11-3/4" L x 5-3/4" W x 5-3/4" D (300 L x 145 W x 145 D)
Tank Capacity: 6 lbs. (2.72 kgs.)
Electrical Requirements:

Inlet Voltage 120 VAC
Frequency 50/60 Hz   
1 Amp
Power Output 0.12 kW

Model PT-18-S Specification:

Overall Dimensions: 21" L x 13" W x 8-1/4" D (535 L x 330 W x 210 D) 
Inner Tank Dimensions: 19-3/4" L x 11-1/2" W x 5-3/4" D (500 L x 290 W x 145 D)
Tank Capacity: 18 lbs. (8.16 kgs.)
Electrical Requirements:

Inlet Voltage 120 VAC
Frequency 50/60 Hz   
3 Amp
Power Output 0.36 kW  

                                                                     

Instructions for Operation and Care of Paraffin Baths

WHITEHALL MANUFACTURING

 • P.O. BOX 3527 • City of Industry, CA 91744-0527 U.S.A

Phone (800) 782-7706 • (626) 968-6681 • Fax (626) 855-4862 • Web: www.whitehallmfg.com

2

Equipment Description

ACCEPTABLE TRANSPORTATION AND 
STORAGE CONDITIONS:

Temperature: +32 ºF to +131 ºF (0 ºC to +55 ºC)
Relative Humidity: 10% to 100%
Atmospheric Pressure: No Requirements

Note: Unit must not be exposed to ingress of water 
or to severe physical trauma. Degradation of unit 
function and/or performance may occur.

CONDITIONS ACCEPTABLES CONCERNANT LE 
TRANSPORT ET L'ENTREPOSAGE:

Température : 

 à +

 (

 à +55 ºC)

Humidité relative : 10 % à 100 %
Pression atmosphérique : aucune exigence

Remarque : l'appareil ne doit pas être exposé à une 
infiltration d'eau ou à un traumatisme physique grave.
Une détérioration des fonctions et/ou du 

rendement de 

l'appareil peut en découler.

+32 ºF

131 ºF 0 ºC

Содержание Para-Therapy PT-18-S

Страница 1: ...ra Therapy Revised April 2017 5900 100 000 INSTRUCTIONS FOR OPERATION AND CARE OF 800 782 7706 626 968 6681 www whitehallmfg com P O Box 3527 City of Industry CA 91744 0527 U S A Division of Acorn Eng...

Страница 2: ...eservoir ou de les appuyercontre celui ci l RETIRER tout emballage accessoire et feuillet se trouvant l int rieur de votre nouvel appareil avant de l utiliser EFFECTUER LA MISE LA TERRE de votre appar...

Страница 3: ...ams and Specifications Model PT 6 S Page 6 Model PT 18 S Page 8 Repair Parts Model PT 6 S Page 6 Model PT 18 S Page 8 Exploded View Model PT 6 S Page 7 Model PT 18 S Page 9 Troubleshooting Guide Page...

Страница 4: ...ns 13 L x 7 W x 7 3 4 D 330 L x 180 W x 195 D Inner Tank Dimensions 11 3 4 L x 5 3 4 W x 5 3 4 D 300 L x 145 W x 145 D Tank Capacity 6 lbs 2 72 kgs Electrical Requirements Inlet Voltage 120 VAC Freque...

Страница 5: ...een pound unit will take 10 to 12 hours to melt all the paraffin completely Once paraffin is completely melted your unit is ready to use Operating Tips When do I need to add Paraffin Add paraffin when...

Страница 6: ...s of the body which cannot be dipped into the tank a soft brush may be used to apply multiple layers of paraffin to area as required for treatment 5 For maximum benefit after the paraffin has been app...

Страница 7: ...towel Carefully lift unit up an inch or two and then bump it down on the surface to break free the solid paraffin block Block should drop on to lid Lift away unit leaving the paraffin block on top of...

Страница 8: ...113 001 Euro Outer Housing 13 6505 540 000 Indicator Light 250 VAC 14 5902 113 199 Bottom Cover 15 6504 604 000 Snap In Bumper Foot 16 0331 014 000 0 203 ID x 0 734 OD Stainless Steel Washer 17 0116...

Страница 9: ...855 4862 Web www whitehallmfg com Instructions for Operation and Care of Paraffin Baths FUSE 1 FUSE 2 GND INLET L N THERMO INDICATOR HEATING ELEMENTS LIGHT STAT THERMAL CUTOUT POWER Electrical Specifi...

Страница 10: ...13 199 Bottom Cover 14 6505 540 000 Amber Indicator Light 125 V 15 0116 010 000 10 32 x 1 2 Phillips Truss Head Screw 16 0302 004 000 10 32 Hex Nut 17 6502 560 000 10 Stainless Steel Lockwasher 6504 6...

Страница 11: ...68 6681 Fax 626 855 4862 Web www whitehallmfg com Instructions for Operation and Care of Paraffin Baths 1 2 3 6 7 23 24 5 3 4 2 9 8 3 10 11 12 13 14 15 16 17 18 19 20 12 25 FUSE 1 FUSE 2 GND INLET L N...

Страница 12: ...MAL CUTOUT BURNT OR FAULTY REPLACE THERMOSTAT THERMOSTAT IS BROKEN HEATING ELEMENT IS REPLACE HEATING ELEMENTS BURNT OR FAULTY PARAFFIN WAX IS MELTING INDICATOR LIGHT REPLACE INDICATOR LIGHT INDICATOR...

Страница 13: ...stitute the sole remedy of the Buyer and the sole remedy of Whitehall under this warranty Whitehall shall not be liable under any circumstances for incidental consequential or direct charges caused by...

Отзывы: