![Whip Mix ProPress SP Скачать руководство пользователя страница 10](http://html.mh-extra.com/html/whip-mix/propress-sp/propress-sp_operators-manual_969659010.webp)
10
ProPress SP
Furnace Operations Manual
Chapter Two — Special Functions
Most Special Functions can be performed while a
program is running, however this is not recommended
because changing the operating features could affect
how the current program runs.
To look at or change a function, press the
Special
Functions
item from the
Main Menu
. Turn the
Selector
Knob
clockwise or counterclockwise until the desired
function is displayed. Press the
Selector Knob
OR the
Next
button to look at or change this function. Each
function will be described in this chapter.
Setting the Night Mode Temperature
and Activation
The furnace has a
Night Mode
feature that will reduce
the temperature in the muffle and raise the lift to prevent
moisture build up.
The
Night Mode
can be started from
idle mode
or while
a program is running.
The furnace will automatically enter
Night Mode
if no
buttons have been pressed for 2 hours, or 45 minutes
have passed since the last run program and no buttons
have been pressed.
The furnace comes from the factory with the
Night Mode
temperature set at 150° C (302° F). This temperature may
be adjusted from 0 to 400° C (752° F).
If
Night Mode
is activated while a program is running, the
furnace will go to the
Night Mode
temperature after the
completion of the program and will maintain the
Night
Mode
temperature until any button is pressed or the
power is turned off.
To adjust the
Night Mode
temperature:
1. From the
Main Menu
, press the
Special Functions
item.
2. Turn the
Selector Knob
until the display reads
NIGHT MODE TEMP.
3. Press the
Next
button to look at or change the night
mode temperature.
4. Turn the
Selector Knob
to change the temperature
as desired.
5. Press the
Next
button to enter the temperature. You
will be returned to the
Main Menu
automatically.
After the
Night Mode
Temperature has been set, follow
these steps to activate
Night Mode
from
idle mode
or
during a program.
1. From the Main Menu, select the
Special Function
item. Turn the
Selector Knob
to the displayed
item
Night Mode.
2a. Push the
Selector Knob
OR press the
Next
button
to activate
Night Mode.
2b. To start
Night Mode
at the end of the program while
a program is running.
Turn the
Selector Knob
to
NEXT
=
NIGHT MODE
and
press the
Selector Knob
or
Next
button.
Change Idle Temperature
The idle temperature is the temperature the furnace will
maintain between programs. The furnace comes from the
factory with the idle temperatures set at 0° C, so you must
enter a temperature for this feature to suit your needs.
1. Press the
Menu
button.
2. Select
Special Functions
, push
Selector Knob
OR
press
Next
button
3. Turn the
Selector Knob
to display
CHANGE IDLE TEMP.
4. Push
Selector Knob
OR press the
Next
button to
look at or change the idle temperature.
5. The current setting will be displayed. Turn the
Selector Knob
to change to the desired temperature.
6. Push the
Selector Knob
OR press the
Next
button
to enter the temperature. You will be returned to the
Main Menu automatically.
NOTE: DURING SHIPPING THE MUFFLE MIGHT ABSORB
MOISTURE FROM THE AIR. SO, IT IS RECOMMENDED
YOU HEAT THE MUFFLE TO DRY THE MOISTURE
BEFORE FIRING ANY PORCELAIN. TO DRY THE MUFFLE,
SET THE IDLE TEMPERATURE TO 400° C (752° F) FOR
ONE HOUR AND THEN 600° C (1112° F) FOR ONE HOUR.
IT IS ALSO RECOMMENDED YOU KEEP THE FURNACE
ON WITH SUFFICIENT NIGHT MODE TEMPERATURE TO
PREVENT MOISTURE FROM ACCUMULATING INSIDE
THE MUFFLE. THE RECOMMENDED TEMPERATURE
IS A MINIMUM OF 150° C (302° F). IF LEAVING THE LAB
FOR AN EXTENDED PERIOD OF TIME, YOU MAY TURN
THE FURNACE OFF WITH THE TABLE IN THE CLOSED
POSITION. BEFORE FIRING PORCELAIN AGAIN, GO
THROUGH THIS DRY OUT PROCEDURE.
*
Special Functions
> Night Mode
Change Idle Temp
Set Vacuum Level