U
SER
M
ANUAL
M
ODEL
M
Page 42
Table 6 Guidance and Manufacturer’s Declaration – Emissions
The Model M is intended for use in the electromagnetic environment specified below. The
customer or user of the Model M should assure that it is used in such an environment.
Emissions Test Compliance
Electromagnetic Environment – Guidance
RF Emissions
CISPR 11
Group 1
The Model M must emit electromagnetic energy in
order to perform its intended function. Nearby
electronic equipment may be affected.
RF Emissions
CISPR 11
Class B
The Model M is suitable for use in all
establishments, including domestic establishments
and those directly connected to the public low
voltage power supply network that supplies
buildings used for domestic purposes.
Harmonics
EN 61000-3-2
Class A
Flicker
EN 61000-3-3
Complies
RF emissions
CISPR 14-1
Not applicable
The Model M is not suitable for interconnection with
other equipment
RF emissions
CISPR 15
Not applicable
The Model M is not suitable for interconnection with
other equipment
Table 7 Guidance and Manufacturer’s Declaration – Immunity
The Model M is intended for use in the electromagnetic environment specified below. The
customer or user of the Model M should assure that it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment –
Guidance
Electrostatic
Discharge (ESD}
IEC 61000-4-2
±6kV Contact
±8kV Air
±6kV Contact
±8kV Air
±8kV Charged
Floors should be wood, concrete or
ceramic tile. If floors are synthetic,
the r/h should be at least 30%
Electrical fast
transient/burst
IEC 61000-4-4
±2kV for power
supply lines
±1kV for
input/output
lines
±2kV for power
supply lines
±1kV for
input/output
lines
Mains power quality should be that of
a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±1kV line(s) to
line(s)
±2kV line(s) to
earth
±1kV line(s) to
line(s)
±2kV line(s) to
earth
Mains power quality should be that of
a typical commercial or hospital
environment.
U
SER
M
ANUAL
M
ODEL
M
Page 19
Figure 7. Pivot the footplate up out of the way
if necessary. Transfer into the chair.
Figure 8. Lower the control arms to the down
position.
Figure 9. If footplate was pivoted up, return it
to the horizontal position.
Figure 10. Turn the chair on then slide the
seat backward using the directional controller.
Recommended practices when transferring into or out of the device include:
●
Position the device so that the distance between it and the object from/to which
you are transferring is close enough for your safe transfer.
●
Use a transfer board if necessary.
●
Put the device in seat mode and slide the seat forward.
●
Turn the device off while transferring.
N026935.01F
320-00630-DHF.DO-01 User Manual Model M US rev 12internal print-paginated-side.pdf
Aug 15 2017 16:01:21
Содержание M
Страница 1: ......
Страница 2: ...USER MANUAL MODEL M 320 00630 DHF DO 01 released 07 17 ...
Страница 13: ...USER MANUAL MODEL M Page 11 3 Product Model M Components Figure 1 Model M Components ...
Страница 61: ...USER MANUAL MODEL M Page 61 ...
Страница 64: ......