PPG Heart Rate Simulator Test System | User Manual
LoadRawData
RawDataFil
ePath
Valid Path,
Valid File
Load raw data from specific
path and file
LoadRawData
D:\Test Data
\case1_replay.txt
PlayRawData
Gain
0.0000000001
~64
Play Raw Data with resize
Gain
times AC part
PlayRawData 0.5
Continue
Duration
Integer
Continue to play
Duration
seconds specific waveform
Continue 8
Reset
N/A
Appear with Play
Raw data
Stop and reset output
Reset
* The detail range for PI, please refer to1.1.2
3.2.4.2
Command Demo
Demo
Explain
MainParameter 8 1.0 60
Play Sine Wave with DC Level=8, PI=1, BPM=60
Continue 17
Continue this wave for 17 seconds
FineTuneParameter 300 20.00 70
Play wave with DC=300 mV, AC=20mV, BPM=70
Continue 6
Continue this wave for 6 seconds
LoadRawData D:\Test Data\case1_replay.txt Load the wave from D:\Test Data\case1_reply.txt
PlayRawData 1.0
Play the case1_reply wave with Gain=1
Continue 20
Continue this wave for 20 seconds
Reset
Stop and reset output signal
MainParameter 8 2.0 60
Play Sine Wave with DC Level=8, PI=2, BPM=60
Continue 25
Continue this wave for 25 seconds
3.3 HRS Software and Firmware Update
HRS unit which were shipped from early 2017 does all support firmware update.
HRS software supports auto-update after software version 1.0.0.6.
If you have any questions, please contact WhaleTeq for detailed information.
3.3.1
Software and Firmware Auto Update
When new firmware or software release, a pop-up window shows as below
–