![WHALETEQ HRS200 Скачать руководство пользователя страница 33](http://html.mh-extra.com/html/whaleteq/hrs200/hrs200_user-manual_967656033.webp)
33
PPG Heart Rate Simulator Test System | User Manual
01
–
Save File
:
Save the completed command list into local PC.
02
–
Load File
:
Load the command file from local PC.
03
–
Execute Command
:
Play and execute the programmed commands.
04
–
Stop Command
:
Stop to execute the ongoing command.
05
–
Loop/Replay
:
Whenever the command has finished executing, re-execute the
command from the beginning.
06 - Help
:
List the command list. Please refer to 3.2.4.1.
07
–
Command List
:
List the programmed or loaded commands. HRS will highlight the
ongoing command.
3.2.4.1
Command Table
Command
Parameter
Range
Description
Example
MainParameter
DC_Level
Integer: 1~20
Send sine wave with specific
DC Level, PI Value and BPM
MainParameter
8 1.0 60
PI
Change with DC
BPM
Integer: 30~300
FineTuneParam
eter
DC_Value
Integer:
100~2500
Send wave with specific DC
Value, AC Value and BPM
FineTuneParameter
300 2000 70
AC_Value
Integer: 0.75~25
BPM
Integer: 30~300