PPG Heart Rate Simulator Test System | User Manual
04
–
Help
:
List all the supported raw-data formats. For details please refer to
section 3.2.3.4.
05
–
Original Waveform Display
:
To display the waveform of loaded raw data. With the use of
03 Time
Interval Selector
, user can quickly check the waveform with preferred
time interval.
06
–
Loop
:
Enable the loops function to allow automatically play the waveform
again once the playback is finished.
07
–
Synchronized
:
Enable the synchronized function to allow HRS LED to response in
accordance with DUT LED.
08
–
Brightness Parameters
:
User can set, save and restore brightness parameters including overall
brightness, gain and offset in statistic and dynamic conditions.
09
–
Time Parameters
:
User can adjust Sample Rate and observe the time frame of current
waveform playback.
10
–
Play and Stop
:
Play and Stop the loaded raw-data waveforms.
11
–
Output Waveform Display
:
Display the current output waveform.
3.2.3.1
Clinical Database Waveform Playback (Optional)
4
To have wearable to perform more closely to medical grade device and
decrease the time and cost of clinical trials, HRS is now built in clinical database
waveforms. All those PPG database waveforms are recorded simultaneously
with ECG signals and have physician to determine the arrhythmia marks via
ECG signals.
HRS clinical database includes: AF (Atrial Fibrillation), APC (Atrial Premature
Complex), VPC (Ventricular Premature Complex) and 1st AV block (First-degree
Atrioventricular block).
Please follow the below instructions to initiate the use:
4
For HRS200 serial number after WHR1602025, please directly download the latest HRS software
(V1.0.0.6) for the trial use of AF database playback function. If HRS200 serial number is before
WHR1602025, please contact WhaleTeq for the upgrade solution.