CMRR 3.0E | User Manual
7
(1)
Place a metal sheet or a metal bench under ECG DUT and CMRR 3.0E
(2)
Connect GND terminal (outer shielding point) of CMRR 3.0E to the metal sheet or the metal
bench.
(3)
Connect the metal frame of EEG DUT to the metal sheet or the metal bench.
(4)
Tester shall keep distance from the test system, to prevent from body capacitance effect. Else,
tester shall touch the metal sheet or the metal bench to connect to the common ground plane.
Figure. 3-1, Set-up CMRR 3.0E Test Environment
3.2
Shielding Covers for the Electrode Cables
(1)
Please fasten the bottom shielding cover to the right-side panel of CMRR 3.0E for reducing main
frequency interference. (see Figure.3-2)
Figure. 3-2, CMRR 3.0E with bottom shielding cover
(2)
As the bottom shielding cover is connected to Vc, Vc would be unstable due to the outer noise.
When Vs is 2V (Vc=1V), Vc has the possibility to exceed 1% tolerance. At the time, please fasten the
top shielding cover with CMRR 3.0E, to control Vc voltage within 1% tolerance. (See Figure. 3-2)