![WHALETEQ CMRR 3.0E Скачать руководство пользователя страница 12](http://html.mh-extra.com/html/whaleteq/cmrr-3-0e/cmrr-3-0e_user-manual_967651012.webp)
CMRR 3.0E | User Manual
12
It connects the DC 12V power supply bundled with CMRR 3.0E to provide the power required by
operation.
5.2.4
Power Switch
It turns on or off the mains supply of DC 12V power supply.
5.2.5
[Vc] Terminal
It connects the common mode point inside CMRR 3.0E to Vc terminal directly. This is intended for use in
equipment calibration.
5.2.6
[Vs] Terminal
The voltage output of CMRR 3.0E built-in sine wave signal generator. This is intended for use in
equipment calibration.
5.2.7
[Monitor] Terminal
The output terminal of inner common mode point after it passes through 11:1 voltage divider. It
measures the voltage of Vc decayed by a factor of 11 directly. This is intended for use in equipment
calibration.
5.2.8
Grounding Terminal
Outer shield grounding that connects the metal sheet in figure 3-1 in tests to reduce noises.
5.3
Right Panel
The right panel of CMRR 3.0E is shown in figure 5-3. It is mainly used to connect each electrodes of EEG.
Figure 5-3, the right panel of CMRR 3.0E