CMRR 3.0E | User Manual
19
Figure 6-7, CMRR 3.0E Assistant Software
Before clicking [Next], if not familiar with the test sequence, user can click [Test Sequence] first to make
the test sequence shown simultaneously, as shown in figure 6-8. The blue-shaded description texts in
[Test Sequence] change as the test steps proceed. This makes the user to have a better understanding of
the test sequences.
Figure 6-8, Software interface and test sequence description of CMRR 3.0E Assistant
6.2.2.2
Step 2. Frequency Setting
Set the line frequency to either 50Hz or 60Hz, or both. When both frequencies are selected, the setting of
50Hz will be performed first, and the same setting would be performed automatically with frequency
replaced with 60Hz.
6.2.2.3
Step 3. Connection
Connect to patient cable. This includes two settings:
1.
Tune Ct until Vc= 1 mV (tune [Coarse]/[Fine] knob)
2.
Connect patient cable to CMRR 3.0E (after tuning Vc)