ECO
FL User’s Guide (FL)
Revision AN
4 June 2012
1
1.
Overview
WET Labs offers a range of the
Environmental Characterization Optics
(
ECO
)
series
fluorometers for the following measurements: chlorophyll, Colored Dissolved Organic Matter,
uranine, rhodamine, and phycocyanin and phycoerythrin. This section includes meter and
connector specifications, delivered items and descriptions of features, including internal batteries
and
Bio-wiper
™.
1.1
Specifications
FL(RT)—
Provides analog or RS-232 serial output with 16,000-count range. This unit provides
continuous operation when powered.
FL(RT)D
—Provides the capabilities of the FL(RT) with 6,000-meter depth rating.
FL—
Provides the capabilities of the FL(RT) with periodic sampling.
FLS
—Provides the capabilities of the FL with an integrated anti-fouling
bio-wiper
™.
FLB
—Provides the capabilities of the FL with internal batteries for autonomous operation.
FLSB
—Provides the capabilities of the FLS with internal batteries for autonomous operation.
FL(RT)
FL(RT)D
FL
FLS
FLB
FLSB
Mechanical
Diameter
6.3 cm
Length
12.7 cm
17.68 cm 12.7 cm
13.3 cm
27.4 cm
27.99 cm
Weight (in air)
0.4 kg
1.3 kg
0.4 kg
0.5 kg
0.96 kg
Weight (in water)
0.02 kg
0.75 kg
0.02 kg
0.08 kg
0.14 kg
Pressure housing
acetal
titanium
Acetal copolymer
Environmental
Temperature range
0–30 deg C
Depth rating
600 m
6000 m
600 m
300 m
Optional pressure sensor
No
Yes
Optional thermistor
No
Yes
Electrical
Digital output resolution
14 bit
Analog output signal
0–5 V
Internal data logging
No
No
Yes
Internal batteries
No
Yes
Connector
MCBH6M
Input voltage
7–15 VDC
Current, typical
50 mA
Current, sleep
--
85 µA
Data memory
--
108,000 samples
Sample rate
to 8 Hz
RS-232 output
19200 baud
Optional anti-fouling
Bio-wiper™
No
Yes
No
Yes
Bio-wiper
™
cycle
--
140 mA
--
140 mA