4. Installation & Operation

4.1 Installation

The water bath must be placed on a smooth, level, and sturdy work surface, preferably near 

a drain for emptying. Use in a ventilated room. Suitable for use in ambient temperatures 5°C 

to 40°C with a maximum humidity 80% (temperature 31°C) decreasing to 50% (temperature 

40°C).

DO NOT block or restrict ventilation slots. DO NOT place directly next to hot heat surfaces.

ENSURE that there is sufficient space around the product to allow it to provide optimum 

temperature control.

When unpacking the unit, remove the packaging and retain over the warranty period.  

Contents consist of:

• 

Water Bath

• 

Stainless Steel False Base

• 

Power Lead

• 

Instruction Manual

Place the false base into the bath with the legs down. Fit the power lead into the socket 

at the rear.

4.2 Operation

Switch the unit on using the mains switch located on the rear. The switch will illuminate,  

and the controller will perform a self-test. When the unit is switched off all time and 

temperature values remain in memory.

Setting Temperature:

1.  Press and hold the DOWN arrow, or the FUNCTION key to display SP1: the set 

temperature. The set temperature indicator will illuminate. 

2.  Use the UP and DOWN arrow keys to select required temperature. 
3.  After setting the temperature the display flashes between SP1 and set temperature 

value and will then revert to show actual liquid temperature. Heater indicator will 

illuminate.

4.  The water bath is now set and will heat and control the liquid to set temperature.

Setting Time:

1.  Press the FUNCTION key until ‘t’ appears on the display. Display will alternate 

between ‘t’ and time displayed as h:mm.

2.  Use the UP and DOWN arrows to select time required. Minimum 0 hours 01 minutes: 

00:01 and maximum 9 hours 59 minutes: 9:59.

3.  Press the FUNCTION key to confirm. The display will flash 4 times then revert to 

actual liquid temperature.

Содержание 663-898

Страница 1: ...1 PRODUCT MANUAL Product Manual Westlab Digital Stirred Water Baths Codes 663 898 663 899 663 900 663 903 663 904...

Страница 2: ...rrer is quiet in use and provides excellent temperature stability Indication of heating over temperature alarm and timer running When a low liquid level is detected a FILL message automatically appear...

Страница 3: ...e Ambient 5 C to 99 C Temperature Alarm Digital LED Display Under Over Temperature Alarm is Automatically Set 4 C Above 4 Below Set Point Temperature Control PID Digital LED Display Resolution of 0 1...

Страница 4: ...he Maintenance section to ensure the proposed method will not damage the unit 8 Connect only to a power supply with the corresponding voltage to that specified on the rating label positioned on the re...

Страница 5: ...within your laboratory ATTENTION Disconnect the Bath from the Power Supply Prior to Cleaning Basic Cleaning Procedure The stainless steel crevice free tank should provide years of valuable service an...

Страница 6: ...alse base into the bath with the legs down Fit the power lead into the socket at the rear 4 2 Operation Switch the unit on using the mains switch located on the rear The switch will illuminate and the...

Страница 7: ...imed period has finished Press the RUN key to deactivate the beep and clear the END message Power Interrupted During Timer Mode If the power is interrupted during the timer mode the display shows P OF...

Отзывы: