Suitable Liquids:

Operating temperatures and recommended liquid options are as follows:

Temperature Range

Options

- 20°C to Ambient

100% Heat Transfer Liquid

100% Car Antifreeze (Ethylene Glycol)

Ambient to + 99°C

Distilled Water

Virkon Dissolved in Distilled Water: Proven efficacy 

against bacteria (including micro bacteria), viruses, 

spores, and fungi in a variety of independent tests 

using different protocols.

Heat Transfer Liquid 

Note: For temperatures above 60°C or below room temperature, the water bath should be covered  
by a lid or layer of polypropylene spheres to achieve optimal performance.

3.2 Maintenance

Please ensure that the washing agent and sanitising agent are BSI accredited and 

approved by the H&S department for use on laboratory equipment and stainless steel 

within your laboratory.

ATTENTION: Disconnect the Bath from the Power Supply Prior to Cleaning  

Basic Cleaning Procedure:

The stainless-steel crevice free tank should provide years of valuable service 

and is resistant to chloride containing solutions, but it is important to avoid high 

concentrations of halogens- particularly chloride. Halogen deposits may show as 

rust which can be cleaned off with nitric acid (10%) on a cloth. WEAR PROTECTIVE 

EQUIPMENT!

The water bath should be emptied at the end of each day, then for interior surfaces:

• WASH • RINSE • DRY •

For algae and slime control in water baths, please use Water Bath Conditioner AR.

Occasionally small spots may occur in the bath which resemble rust. These are usually 

small ferrous particles which have oxidised. Using a stainless-steel cleanser such as 

Bar Keepers Friend® will remove these marks as well as mineral deposits.  

Exterior Anti-Bacterial Painted Surfaces:

The water bath should be cleaned at regular intervals by wiping external surfaces with 

a cloth or sponge soaked in warm water with a mild detergent.  

ATTENTION: Do Not Use Strong Solvents or Solutions Containing Chlorinated 

Hydrocarbons, Esters, Ketones or Abrasive Cleaners as this may Damage the Built-In 

Antibacterial Properties.

Содержание 663-898

Страница 1: ...1 PRODUCT MANUAL Product Manual Westlab Digital Stirred Water Baths Codes 663 898 663 899 663 900 663 903 663 904...

Страница 2: ...rrer is quiet in use and provides excellent temperature stability Indication of heating over temperature alarm and timer running When a low liquid level is detected a FILL message automatically appear...

Страница 3: ...e Ambient 5 C to 99 C Temperature Alarm Digital LED Display Under Over Temperature Alarm is Automatically Set 4 C Above 4 Below Set Point Temperature Control PID Digital LED Display Resolution of 0 1...

Страница 4: ...he Maintenance section to ensure the proposed method will not damage the unit 8 Connect only to a power supply with the corresponding voltage to that specified on the rating label positioned on the re...

Страница 5: ...within your laboratory ATTENTION Disconnect the Bath from the Power Supply Prior to Cleaning Basic Cleaning Procedure The stainless steel crevice free tank should provide years of valuable service an...

Страница 6: ...alse base into the bath with the legs down Fit the power lead into the socket at the rear 4 2 Operation Switch the unit on using the mains switch located on the rear The switch will illuminate and the...

Страница 7: ...imed period has finished Press the RUN key to deactivate the beep and clear the END message Power Interrupted During Timer Mode If the power is interrupted during the timer mode the display shows P OF...

Отзывы: