background image

3

PRODUCT MANUAL

1800 358 101

[email protected]

www.westlab.com.au

+64 9 553 3677

[email protected]

www.westlab.co.nz

646 568 5391

[email protected]

www.westlab.com

1 877 822 1455

[email protected]

www.westlab.com

Australia

New Zealand

United States

Canada

2. Technical Specification

663-898

663-899

663-900

Tank Capacity

14L

22L

28L

Temperature 

Range

A 5°C to 99°C

Temperature 

Alarm

Digital LED Display - Under/Over Temperature Alarm is 

Automatically Set 4°C Above & 4° Below Set Point

Temperature 

Control

PID Digital, LED Display, Resolution of 0.1°C

Water Level Range

130mm Maximum

Sensitivity

± 0.05°C

Uniformity

± 0.01°C

Timer

Visual Indicator & End Message  

– Variable 00:01 to 99:59 hh:mm

Heater Power

1,250W

Voltage

230 V, 50 to 60 Hz

Internal 

Dimensions (mm)

219 W x 298 D  

x 150 H

395 W x 298 D  

x 150 H

395 W x 298 D  

x 200 H

External 

Dimensions (mm)

361 W x 332 D  

x 170 H

537 W x 332 D  

x 170 H

537 W x 332 D 

 x 290 H

Содержание 663-898

Страница 1: ...1 PRODUCT MANUAL Product Manual Westlab Digital Stirred Water Baths Codes 663 898 663 899 663 900 663 903 663 904...

Страница 2: ...rrer is quiet in use and provides excellent temperature stability Indication of heating over temperature alarm and timer running When a low liquid level is detected a FILL message automatically appear...

Страница 3: ...e Ambient 5 C to 99 C Temperature Alarm Digital LED Display Under Over Temperature Alarm is Automatically Set 4 C Above 4 Below Set Point Temperature Control PID Digital LED Display Resolution of 0 1...

Страница 4: ...he Maintenance section to ensure the proposed method will not damage the unit 8 Connect only to a power supply with the corresponding voltage to that specified on the rating label positioned on the re...

Страница 5: ...within your laboratory ATTENTION Disconnect the Bath from the Power Supply Prior to Cleaning Basic Cleaning Procedure The stainless steel crevice free tank should provide years of valuable service an...

Страница 6: ...alse base into the bath with the legs down Fit the power lead into the socket at the rear 4 2 Operation Switch the unit on using the mains switch located on the rear The switch will illuminate and the...

Страница 7: ...imed period has finished Press the RUN key to deactivate the beep and clear the END message Power Interrupted During Timer Mode If the power is interrupted during the timer mode the display shows P OF...

Отзывы: