25
Temperature Sensor for Contactless Temperature Measurement
7.17. rESP (set response time)
Setting
Description
Value in s
Response time can be set to one of the following values:
0.065 s, 0.10 s, 0.34 s, 1.10 s, 1.33 s, 3.00 s, 5.00 s, 10.00 s, 30.00 s
The higher the response time setting, the greater the extent to which temperature
peaks and fluctuations can be filtered. Adjusting response time changes the switching
frequency.
Note!
Switching frequency is reduced as response time is increased.
7.18. SEr.O (activate or deactivate the continuous temperature read-out)
The sensor is equipped with an RS 232 port for communication with a device such as
a PC or a controller.
All sensor settings can be selected digitally with a PC, and all values generated by the
sensor can be read out at a PC.
Interface configuration:
Baud rate: 38,400 baud, COM: 34800,n,8,1
Setting
Description
On
Continuous temperature read-out On
The sensor reads out the measured temperature once every 100 ms.
Note!
Configuration via wTeach software is not possible in this setting.
OFF
Continuous temperature read-out Off
The sensor can be configured and/or queried via the serial port.
All read-outs, including temperature, are executed as the result of an RS 232
command.
Note!
If the sensor will be set up via the serial port (wTeach software), this setting
must be selected.