Safety instructions and warnings
01/2022
1020143-00
Operating instructions – 13/52
Setup location
WARNING!
Overheating from inadequate cooling!
The unit might be damaged!
–
Maintain the minimum distances from the walls! Dimen-
sions are given on page 44.
–
Do not change, obstruct or block the air inflow and
outflow to the unit or make any unauthorized changes
to the unit.
Salty air! Contaminated air!
Unit may be damaged due to corrosion of the housing and
of electric components.
–
Where possible, do not set up the unit at a swimming
pool.
–
Do not operate the unit in locations with aggressive
ambient conditions (for example, with air containing
chlorine, in an oxygen enriched environment).
Damage to unit possible!
Where there are great differences in temperature between
the transport route and the installation site the unit may not
be put into operation immediately after being set up.
–
Wait at least 2 hours before connecting the unit to the
power supply.
Environmental conditions
Optimum ambient temperature: B15 °C and +32 °C
Storage and transport temperature:
Storage temperature: Between -15 °C and +60 °C (the well, pump and
the hoses must not contain any water)
Relative humidity and air pressure for transport, storage and operation:
Relative humidity: Between 30 % and 75 %
Air pressure: Between 794 hPa and 1060 hPa
Water and cooling conditions
Permissible water pressure (water inlet): 0.8 MPa (8 bar)
Maximum coolant temperature: 16°C
The unit is not to be used outdoors.
The unit is not suitable for operation in mobile facilities (ships, buses, trains or
oil platform). The unit has to be adapted in order to guarantee trouble-free
operation when used in mobile facilities.
Adaptation of the unit is required for operating altitudes greater than 2000 m
above sea level in order to guarantee trouble-free operation.
Please contact the Customer service department without fail before putting
the unit into operation - see page 2.