15
WELLION
L
UNA
DUO
CONTROL SOLUTION TEST
The control solution is used to check the performance of the Wellion LUNA duo blood
glucose and total cholesterol measuring system. The system and strips perform ade-
quately if the control test result falls within the indicated control range listed on the
test strip vial.
Use only the Wellion LUNA Control Solutions with the Wellion LUNA Test Strips. Other
brands of control solution will produce inaccurate results.
The system should be checked:
• When you get your meter for the first time or start using a new box of test strips.
• When you suspect that the meter or the test strips are not working properly.
• Any time you get unexpected test results.
• If you drop the meter.
• If you want to check that your testing procedure is correct.
1. Fully insert the contact bar of the test strip into the test port of the meter, with the
labelled side of the strip facing upwards. Do not bend the test strip.
The respective test mode will be shown on the screen depending on which test strip
you‘re using. After the strip has been inserted, the blood drop icon will blink on the
screen which means the system is ready for blood to be applied to the strip.
Содержание LUNA duo