57
E 3
Test strip
removed during
countdown
Repeat the test with a new test strip. See the
chapters “Testing Your Blood Glucose“ or “Testing
Your Blood Cholesterol“
E 4
Meter detects
extreme condition
Move the meter into an area that is between 5°C
to 45°C for glucose testing or 18°C to 38°C for
cholesterol testing and allow 10 to 15 minutes for
it to reach the new temperature. See the chapter
“Storage and Precautions”
E 5
Meter failure
• Replace the batteries and turn the meter on
again.
• Remove the test strip and insert a new test strip.
• If problem persists, contact your local distributor.
E 6
Expired
cholesterol test
strip
• Check if the date and time on your meter is set up.
See “Setting the Clock“
• Verify the expiration date of the cholesterol test
strips
• Verify that the code number on the code card mat-
ches the code number on the test strip package.