
24
CONTROL SOLUTION TESTING
Performing a Glucose Control Solution Test
The purpose of glucose control solution testing is to validate the performance of the Wellion
LEONARDO GLU/CHOL (plus) Blood Glucose and Cholesterol Monitoring System using the
testing solution with a known range of glucose. Control solutions are sold separately.
You should perform control solution testing when:
• Using the meter for the first time
• You open a new bottle of Wellion LEONARDO Blood Glucose Test Strips
• You leave the cap of the test strip vial open for a while
• You drop the meter
• You suspect your Wellion LEONARDO GLU/CHOL (plus) Meter and Wellion
LEONARDO Blood Glucose Test Strips are not working properly
• The blood glucose test results do not reflect how you feel
• You want to practise the testing procedure
Materials you will need:
Wellion LEONARDO Glucose Control Solution (Level 0, Level 1 or Level 2)
Your Wellion LEONARDO GLU/CHOL (plus) Meter
A new Wellion LEONARDO Blood Glucose Test Strip
Содержание LEONARDO GLU/CHOL plus
Страница 1: ...OWNER S GUIDE ...
Страница 23: ...23 ...