28
C
ONTROL
SOL
UTION
TESTING
Step 2:
The meter will light up and show a test strip and flashing blood
drop icon. If other messages (E 1, E 2, etc) appear, see Troubleshoo-
ting on page 30.
Step 3:
Squeeze a drop of Wellion CALLA Control Solution onto a clean,
dry, nonabsorbent surface. Do not apply control solution to the test strip directly
from the bottle. Replace the bottle cap on the control solution bottle immediately
after use.
Step 4:
While holding the meter, touch the control solution to the edge
of the test strip, and the control solution would be automatically
sipped into the reaction area of the test strip. Hold the meter until
it beeps.
• Make sure you are testing between 10°C - 40°C. If the Meter displays ,
move the meter into an area that is between 10°C - 40°C and allow 10 to
15 minutes for it to reach the new temperature before use. Your Meter will not
begin testing if it’s outside this temperature range.