C
ONTROL
SOL
UTION
TESTING
Step 5:
The screen will count down. After 6 seconds, the control solution testing result will
appear on the meter display screen.
Step 6:
Compare the reading on the screen to the corresponding con-
trol range printed on the test strip bottle. If the reading does
not fall within the ranges printed on the test strip bottle, see
Control Solution Solving Problems on page 45.
Step 7:
Remove and discard the used test strip into proper waste bas-
ket. The meter will automatically shut off after the test strip has been removed, or
will shut off after 2 minutes of nonuse.
Perform other levels of control solution testing by repeating the above steps.
NOTE:
• Control solution testing results will be stored in the meter memory and indi-
cated by ctl icon.
• Control solution testing result will not be used for calculating averages.
• DO NOT reuse test strips.