4. To verify that the file was imported successfully, screen a new subject, or select a
completed screening and print the report. Your new banner prints on the bottom of
the report.
Web access
When connected to a wireless network, the vision screener gives you the ability to view
files from a web browser (for example, Internet Explorer, Firefox, Safari) on a computer
on the same network.
To access the webserver on the vision screener
Make sure your computer and the vision screener are on the same network.
1. From your computer, open a web browser.
2. Enter the vision screener IP address in the address bar, for example
123.45.67.89
. The IP address is located on the top black bar on the Home screen.
3. If the web browser displays a security warning, proceed.
4. Enter the username and password.
Username:
spot
Password:
0000
Note
If you have enabled security on your device, the password is
your four-digit pin code.
5. To view the exam results of each individual screening, select the pdf folder.
Vision screener criteria
Spot Vision Screener’s criteria was developed in accord with the American Academy of
Pediatric Ophthalmology and Strabismus (AAPOS) and American Academy of Pediatrics
(AAP) recommendations for instrument based vision screening. The criteria is age
specific to assist with early detection of amblyopic risk factors in children.
WARNING
Risk of false positive and false negative results. Modification of
referral criteria should be performed only by a vision specialist.
Note
Any deviation from this criteria may lead to over referrals or under referrals.
If you enter criteria that is not within accepted limits, the vision screener
automatically restores the system default criteria.
Note
Verify the system clock settings after modifying criteria.
•
The subject age in months shall have no overlapping ages between the age ranges.
•
No subject age range can span 240 months (20 years).
•
Anisometropia Threshold Value(age): Range 0.25D to 14.00D in increments of 0.25
diopters.
•
Myopia Threshold Value(age): Range 0.00 to +7.00 diopters in increments of 0.25
diopters.
34 Advanced settings on the vision screener
Welch Allyn
®
Spot
™
Vision Screener Model VS100
Содержание Spot VS100
Страница 1: ...Welch Allyn Spot Vision Screener Model VS100 Directions For Use Software version 3 1 XX ...
Страница 18: ...14 Controls display window and connections Welch Allyn Spot Vision Screener Model VS100 ...
Страница 22: ...18 Setting up the vision screener Welch Allyn Spot Vision Screener Model VS100 ...
Страница 58: ...54 Specifications Welch Allyn Spot Vision Screener Model VS100 ...
Страница 60: ...56 Warranty Welch Allyn Spot Vision Screener Model VS100 ...
Страница 62: ...58 Service policy Welch Allyn Spot Vision Screener Model VS100 ...
Страница 64: ...Material No 728900 ...