Preliminary Bench Test for MicroTymp 2 23641 continued:
Test 5 Inactive Current Drain
5.1 — Keep battery in partially installed position as in step 4.2 above.
5.2 — Connect Milliammeter ( - ) Negative lead to positive spring contact in the battery
compartment as per FIG 5.2
5.3 — Connect Milliammeter (+) Positive lead to positive side of battery . FIG 5.2
5.4 — Observe current drain. Current will be less than 1 milliamp, typically 500-600
microamps. Note: One possible cause of failure is shorted capacitor C11.
5.5 — Leave all connections as they are for the next test.
Test 6 Active Current Drain
6.1 — Depress and hold black center push button to activate MT2.
6.2 — Observe current drain while the black button is depressed. Current can rise up
to as much as 350 mA +/- 50 mA , then fall to 175 mA +/- 25 mA..
6.3 — Release black button. Current will fall to 135 mA +/- 25 mA.
6.4 — If readings are normal, calibrate with Calvin equipment.
4
Figure 5.2 Current Drain
Probe end
Valve
+ Milliamps –
battery
+
–
+
Inactive current drain 500 600 microamps
Button IN Active current drain 175 +/- 25 mA
Button OUT Active Current drain 135 +/- 25mA
Содержание Micro Tymp 2
Страница 55: ...MT2 DRAWINGS NOS 236300 236302 236304 236305 236308 236330 236340 711700 MICRO TYMP 2 HARDWARE BLOCK DIAGRAM...
Страница 56: ......
Страница 57: ......
Страница 58: ......
Страница 59: ......
Страница 60: ......
Страница 61: ......
Страница 62: ......
Страница 63: ......
Страница 64: ......
Страница 65: ...SECTION 11 A 02727 PROCESS SHEET EXCERPTS FOR 23641 MICROTYMP 2 HANDLE TESTS...
Страница 66: ...SECTION 12 A 02728 Process Sheet excerpts for 71171 71172 71174 71176 71179 Micro Tymp 2 Printer Charger tests...