C

LEANING

L

ENSES

• Remove the focusing sleeve from the fiber optic bundle by unthreading counter-

clockwise.              

• Use a cotton swab soaked in warm water and mild detergent to clean both the  

distal end of the fiber optics and the proximal side of the small lens. (See diagram 
for locations.)

• Reassemble focusing sleeve to Light Pipe.

• The window may also be cleaned with warm water and mild detergent. 

L

IGHT

P

IPE

• To maintain unit cleanliness, use disposable Sheath (No. 52640).

• May be cleaned with a mild detergent and water solution.

W

ARNING

:

Do not clean lens or window with alcohol.

L

IGHT

B

OX

Do not Sterilize.
Do not Immerse in Cleaning Solutions.

1.

Unplug the unit from the wall outlet prior to cleaning.

2.

Insert the red plastic plug into the Light Pipe socket 
prior to cleaning. This will prevent the cleaning solution 
from getting into the filter inside the unit.

3.

Wipe the surfaces of the unit with a cloth dampened in 
a mild solution of detergent and water. Wipe dry with a 
clean, dry cloth.

4.

The power cord may be wiped clean using the same detergent solution as the 
box. Care must be taken not to get the plug prongs wet. Wipe dry with a clean, 
dry cloth.

5.

Do not plug the unit back into the wall outlet until it is thoroughly dry.

6

Light Pipe

Small Lens

Focusing
Sleeve

Distal End 
Of Fiber Optics

Cotton Swab

Содержание Halogen Exam Light II 48600

Страница 1: ...trouble free service WARNING The user of this equipment should be thoroughly trained in the medical proce dures appropriate to the instrumentation Furthermore time should be taken to read and understa...

Страница 2: ...all while pushing down to lock screw heads into small end of key hole slots TABLE MOUNTING NOTE Some table manufacturers supply table mounting hardware for mounting the Exam Light II Check with your t...

Страница 3: ...N NOTE Please refer to use of light pipe section prior to using unit To ensure long unit life high light out put and continued position rigidity we recommend that you do not acutely bend the Light Pip...

Страница 4: ...the desired position and then lower to the desired position SPECIAL NOTE Upon manufacture the Light Pipe is tested to assure maintenance of position and absence of noise Over a period of time it is no...

Страница 5: ...ram 2 Grasp expired lamp by its base and remove from lamp holder 4 IMPORTANT Reposition release lever up to original position prior to installing new lamp Replace only with a Welch Allyn Lamp No 04200...

Страница 6: ...dow with alcohol LIGHT BOX Do not Sterilize Do not Immerse in Cleaning Solutions 1 Unplug the unit from the wall outlet prior to cleaning 2 Insert the red plastic plug into the Light Pipe socket prior...

Страница 7: ...00 120 VAC 60 Hz 0 35A 12 VAC 3 0A 500 MA 48620 100 VAC 50 60 Hz 0 42A 12 VAC 3 0A 630 MA 48622 220 VAC 50 Hz 0 19A 12 VAC 3 0A 200 MA 48624 240 VAC 50 Hz 0 18A 12 VAC 3 0A 200 MA 48626 240 VAC 50 Hz...

Страница 8: ...drift once positioned if mounted vertically according to the enclosed instructions This warranty is for a two year period from original date of purchase provided the Light Pipe has not been tampered w...

Отзывы: