background image

M

OUNTING

W

ALL

M

OUNTING

• Choose appropriate mounting location.

• Using mounting plate as template, mark drilling holes as needed for your wall type:

Dry wall, paneling, thin plywood or plaster 
walls less than 1 inch (2.5 cm) thick. 
Drill four .31 inch (.79 cm) holes, insert hollow 
metal wall anchors making sure barbs are extended.

Concrete block or thick plaster walls 
greater than 1 inch (2.5 cm) thick. 
Drill four .25 inch (.64 cm) holes, insert plastic anchors.

Metal panel or wood walls greater 
than 1 inch (2.5 cm) thick. 
Drill four .31 inch (.79 cm) holes.

• Install mounting plate with appropriate screws. When tightened, there will be a gap 

between the screw heads and surface of the mounting plate.

• Mount unit by fitting key hole slots on back of unit over screw heads.

• Push unit against wall while pushing down to lock screw heads into small end of key 

hole slots.

T

ABLE

M

OUNTING

N

OTE

Some table manufacturers supply table mounting hardware for mounting the

Exam Light II. Check with your table supplier for availability.

• Determine side of table for mounting the Exam Light II.

• Trace the outline of the mounting plate and mark the hole locations on a 

1.00 inch (2.5 cm) thick piece of wood.

• Cut the piece of wood to the outline size and drill four .25 inch (.64 cm) holes.

• Using the mounting plate as template, mark the hole locations on the table and 

drill four .25 inch (.64 cm) holes.

N

OTE

Be sure mounting location is at least 10 inches (25.4 cm) below table cushion,

and that mounting will not obstruct table interior.

• Using four 10-32 x 2 inch (5.08 cm) machine screws with lock washers and nuts, 

install the block of wood and the mounting plate to the table as shown in diagram.

• Mount unit as with wall mounting.

2

Wood Block

Mounting Plate

2 Inch (5.08 cm)
Long Screws

Содержание Halogen Exam Light II 48600

Страница 1: ...trouble free service WARNING The user of this equipment should be thoroughly trained in the medical proce dures appropriate to the instrumentation Furthermore time should be taken to read and understa...

Страница 2: ...all while pushing down to lock screw heads into small end of key hole slots TABLE MOUNTING NOTE Some table manufacturers supply table mounting hardware for mounting the Exam Light II Check with your t...

Страница 3: ...N NOTE Please refer to use of light pipe section prior to using unit To ensure long unit life high light out put and continued position rigidity we recommend that you do not acutely bend the Light Pip...

Страница 4: ...the desired position and then lower to the desired position SPECIAL NOTE Upon manufacture the Light Pipe is tested to assure maintenance of position and absence of noise Over a period of time it is no...

Страница 5: ...ram 2 Grasp expired lamp by its base and remove from lamp holder 4 IMPORTANT Reposition release lever up to original position prior to installing new lamp Replace only with a Welch Allyn Lamp No 04200...

Страница 6: ...dow with alcohol LIGHT BOX Do not Sterilize Do not Immerse in Cleaning Solutions 1 Unplug the unit from the wall outlet prior to cleaning 2 Insert the red plastic plug into the Light Pipe socket prior...

Страница 7: ...00 120 VAC 60 Hz 0 35A 12 VAC 3 0A 500 MA 48620 100 VAC 50 60 Hz 0 42A 12 VAC 3 0A 630 MA 48622 220 VAC 50 Hz 0 19A 12 VAC 3 0A 200 MA 48624 240 VAC 50 Hz 0 18A 12 VAC 3 0A 200 MA 48626 240 VAC 50 Hz...

Страница 8: ...drift once positioned if mounted vertically according to the enclosed instructions This warranty is for a two year period from original date of purchase provided the Light Pipe has not been tampered w...

Отзывы: