N
EONATE
T
RANSILLUMINATOR
C
AUTION
:
Use only Welch Allyn Transilluminator No. 48210 for neonatal transillumination.
S
ET
U
P
A
ND
O
PERATION
Refer to Set Up page for installation steps.
L
AMP
R
EPLACEMENT
W
ARNING
:
Lamps operate at a high temperature and under pressure may shatter. Do not attempt
to replace lamp while hot. Wait until lamp is cool.
1. Turn power switch to “Off” position, unplug cord from outlet and from Light box.
2. Release interlock by sliding interlock arm to left. Pull lamp holder until lamp is fully
exposed. (Lamp holder will not slide down if Interlock is not released)
Diagram #1.
3. Push release lever down; Diagram #2. Grasp expired lamp by its base and remove
from lamp holder.
4. IMPORTANT: Reposition release lever up to original position prior to installing new
lamp. Replace only with a Welch Allyn Lamp No. 04200.
5. Visually inspect replacement lamp reflector and bulb, and remove any grease or
fingerprints with a clean cloth moistened with alcohol, if required.
6. Slide new lamp into lamp holder socket until it “snaps” into place. Make sure
lamp is securely seated to assure maximum light output.
7. Slide lamp holder back into light box. Reengage interlock by moving arm to
the right.
W
ARNING
:
When sliding the lamp holder back into the light box, hold the light box down to avoid
pushing it off of the mounting plate and possibly dropping it.
8. Plug line cord back into unit, and into wall outlet.
5
Diagram #1
Diagram #2
Interlock Arm