Technical Manual
Model 600
29
SCREW,
THD-ROLL #4X1/4 P-PL-S-Z
83021-0000
4
SCREW, PLASTITE, 4-20 X 7/8 FLAT
HEAD 82 DEG PL-S-Z
83035-000
2
SCREW, MACH 4-40 X 1 1/16 P-PL-S-Z
83036-0000
2
SCREW, PLASTITE 4-20 X 3/4 - FLAT
HEAD 82 DEG PL-S-Z
83037-000
4
MODEL 600 PREVENTATIVE MAINTENANCE
The following suggested preventative maintenance is recommended to maximize uninterrupted service
with the Model 600 Thermometer.
Units which are in service on a regular basis should have the following preventative maintenance
performed every 18 months:
1. Visually inspect the thermometer and probe for physical damage which might cause future product
failure.
2. Clean unit if necessary and replace batteries according to the procedures found on page 22 of this
manual.
3. Perform the Power Up Display Test, Startup Display Test and Calibration Key Procedure found on
page 21 and 22 of the manual.
Units which are stored for extended period and not used should have the following preventative
maintenance performed every 36 months:
1. Replace batteries according to the procedures found on page 22 of this manual.
2. Perform the Power Up Display Test, Startup Display Test, and Calibration Plug Procedure found on
page 22 of this manual.
Содержание Diatek 600
Страница 1: ...Diatek Model 600 Thermometer Technical Manual Thermo Menu Operation Manual...
Страница 13: ...Technical Manual Model 600 11 Figure 1 Block Diagram...
Страница 14: ...Diatek A Welch Allyn Company 12 Model 600 Figure 2 Schematic...
Страница 15: ...Technical Manual Model 600 13 Figure 3 Schematic earlier serial numbers...
Страница 16: ...Diatek A Welch Allyn Company 14 Model 600 Figure 4 Printed Circuit Board Component Layout...
Страница 17: ...Technical Manual Model 600 15 Figure 5 Final Assembly...
Страница 18: ...Diatek A Welch Allyn Company 16 Model 600...
Страница 19: ...Technical Manual Model 600 17 Figure 6 Operating Controls and Components...
Страница 32: ...7420 Carroll Road San Diego California 92121 800 854 2904 619 621 6600 FAX 619 621 6610 70017 000H...