Welch Allyn Diatek 600 Скачать руководство пользователя страница 10

Diatek, A Welch Allyn Company

Model 600

2. The low battery detector output (Ul-9) is tested by the microprocessor. A low logic level on Ul-9

indicates a low battery condition (less than 3.0 volts).

3. The horn drive signal (U2-11) goes high activating the horn test for 0.1 second.

4. The backlight switch, normal/monitor switch and F/C switches are tested by the microprocessor.

5. A calibration cycle is performed.

6. The calibration test is performed.

7. The microprocessor memory test is performed.

8. The probe resistance is tested for an open or shorted condition.

Following the events which take place during the display test, a continuous monitor mode temperature
measurement or a predict mode temperature taking cycle is started.

Probe Temperature Measurement

Q3-Q6, Q8, R3-R5, R-16, C2 and U1 comprise a unique resistance to pulse width conversion circuit
which allows any one of 4 resistance’s to be measured by the microprocessor. The circuit allows the
microprocessor to select precision resistors with equivalent temperatures at opposite ends of the
temperature measurement range and measure their relative pulse widths.

The microprocessor sets U2- 16 to 19 high as required to select the appropriate resistance to be
measured. R3 provides a pulse width calibration at 93.2 °F, R5 provides calibration at 106.9 °F and R3
is used for a calibration test at 100.5 °F. A ratiometeric calculation is used to compute the probe
resistance from its pulse width relative to those provided by R3 and R5. The accuracy of each
calibration resistor is .05 °F. A resistance to pulse width conversion is initiated by the microprocessor
setting U2-64 high for approximately 0.5 seconds with the appropriate FET switch selected sometime
within that period. U2-64 is then set low and the time from this transition until the A/D output (U1-2)
goes high is measured by the microprocessor. The crystal X1 provides the time base for the pulse
width measurements. The probe temperature is computed using the probe resistance value obtained
from the ratiometeric pulse width calculations.

Monitor Mode Operation

In the monitor mode a new probe temperature measurement is completed every 1.5 seconds and a
new calibration cycle is completed every 3 seconds. The computed probe temperature is displayed
directly on the liquid crystal display and will follow both upward and downward movement.

Normal Mode Operation

In the normal mode. a new probe temperature measurement is completed every 1.5 seconds. At least
one calibration cycle is performed at the start of a predict cycle and continue every 3 seconds until the
probe temperature rises above 84.0 °F. The calibration cycle is stopped at this point to reduce
quantizing errors associated with the measurement of the slowly changing probe temperature.

The normal mode displayed temperature is the sum of the actual probe temperature and a computed
correction factor which is based on the shape of the changing temperature curve at that point in time.
The values of the correction factor can vary from 0 °F to 2.3 °F with a typical value when the horn
sounds of about 1 °F. During the display test prior to starting a normal mode temperature cycle, the
ambient temperature of the probe is measured and is used in the correction factor computation to
adjust for differing probe starting conditions. The normal mode display is peak reading and therefore
the highest temperature is displayed even though the present predicted probe temperature may drop.

Содержание Diatek 600

Страница 1: ...Diatek Model 600 Thermometer Technical Manual Thermo Menu Operation Manual...

Страница 2: ...reserved No part of this manual may be reproduced or transmitted in any form or by any means electronic or mechanical including photocopy without prior consent in writing from Diatek Printed in the U...

Страница 3: ...st 6 Battery Life 6 Shelf Life 6 Temperature Taking Without Backlight 6 Temperature Taking With Backlight 6 C F Mode Switch 6 Normal Monitor Mode Switch 6 MODEL 600 THEORY OF OPERATION 7 Technical Ove...

Страница 4: ...alibration Check 21 Microprocessor Self Check 22 Cleaning and Sterilization 22 Battery Replacement 22 Calibration Key Procedure 22 Test Procedure 23 M9600 Procedure 24 MODEL 600 TROUBLESHOOTING 24 PAR...

Страница 5: ...uct and the item should be immediately returned securely packaged and postage prepaid to Diatek Loss or damage in shipment shall be at purchaser s risk Diatek will not be responsible for loss associat...

Страница 6: ...horn beeping twice every 10 seconds and the up arrow blinking Five minutes of this error condition causes the thermometer to turn off Returning the probe to its holder and then removing it will restar...

Страница 7: ...ess will abort at that time The low battery error displays a flashing symbol with a double horn beep sounding every 10 seconds The unit will shut off automatically after five minutes of error indicati...

Страница 8: ...number of temperatures that can be taken should be about 2 000 hours 30 temperatures hour 60 000 temperatures Temperature Taking With Backlight The backlight uses about 20 times more power than the re...

Страница 9: ...n and calibration resistor selection To extend battery life the power to both the analog lC U1 and the LCD reference are turned off and the microprocessor is put to sleep when possible Power to the th...

Страница 10: ...or approximately 0 5 seconds with the appropriate FET switch selected sometime within that period U2 64 is then set low and the time from this transition until the A D output U1 2 goes high is measure...

Страница 11: ...n error is displayed Broken Probe Test Each probe resistance sample that is taken is checked to see that it lies between 2000 and 60 000 ohms A resistance value outside of these limits will cause a br...

Страница 12: ...ich in turn causes the low battery indication to be displayed Pulse Timer Operation Pressing the pulse timer switch activates the microprocessor interrupt to U2 63 starting the pulse timer display The...

Страница 13: ...Technical Manual Model 600 11 Figure 1 Block Diagram...

Страница 14: ...Diatek A Welch Allyn Company 12 Model 600 Figure 2 Schematic...

Страница 15: ...Technical Manual Model 600 13 Figure 3 Schematic earlier serial numbers...

Страница 16: ...Diatek A Welch Allyn Company 14 Model 600 Figure 4 Printed Circuit Board Component Layout...

Страница 17: ...Technical Manual Model 600 15 Figure 5 Final Assembly...

Страница 18: ...Diatek A Welch Allyn Company 16 Model 600...

Страница 19: ...Technical Manual Model 600 17 Figure 6 Operating Controls and Components...

Страница 20: ...able batteries three type AA cells up to 1000 hours of continuous use Dimensions length 6 25 inches 159mm width 3 7 inches 94mm height 2 4 inches 61mm weight 10 5 ounces 298 grams case material ABS Cy...

Страница 21: ...he normal monitor slide switch to the normal position Withdraw the probe from the storage channel and observe the 6 second display test ensuring that no display segments are missing The unit will then...

Страница 22: ...the thermometer causing the 6 second display test to appear Verify that no display symbols are missing and then attach the probe to the patient as required Once the temperature readout has been allowe...

Страница 23: ...ery condition exists will cause a predict temperature in progress to be terminated Probe Position Error The probe position error is an indicator that appears whenever there is a rapid drop in temperat...

Страница 24: ...erilization Battery Replacement Remove the battery access screw by turning it counter clockwise with a Phillips screwdriver Slide the battery access cover away from the battery access label to expose...

Страница 25: ...led activate the F C backlight and timer switches and observe that the proper response is obtained Insert and remove a probe to test the probe switch action Placing the normal monitor switch in the no...

Страница 26: ...use of the M9600 refer to the Troubleshooting section in the M9600 Operation Manual MODEL 600 TROUBLESHOOTING GENERAL SYMPTOM PROBLEM TREATMENT No Display Temp or Timer or low Battery Indication Batt...

Страница 27: ...volts Off Cal Check Q3 Q4 Q5 and Q6 as above If the above tests fail to show the failed component replace C2 then U1 Missing or Faded LCD Segments 1 Loose LCD frame 2 Dirt or oil under zebra strip 3 E...

Страница 28: ...S 5760 1 8W 1 RN50 40265 761 R5 RES 4990 1 8W 1 RN50 40264 9910 R6 RES 200K 1 4W 5 40040 4730 R7 RES 33K 1 4W 5 40040 3340 R8 R9 R12 RES 100K 1 4W 5 40014 000 R10 RES 4 3 OHM 1 4W 5 40040 430 R11 RES...

Страница 29: ...FET PWR VN10KM 50021 0000 R1 R2 RES 18K 1 4W 5 40040 1840 R3 RES 6810 1 8W 1 RN50 40266 811 R4 RES 5760 1 8W 1 RN50 40265 761 R5 RES 4990 1 8W 1 RN50 40264 9910 R6 RES 200K 1 4W 5 40040 2050 R7 R8 R9...

Страница 30: ...4 Q1 Q2 Q7 TRANSISTOR 2N2222 50004 0000 4 Q8 TRANSISTOR 2N4124 50014 0000 2 Q3 Q4 Q5 Q6 FET PWR VN10KM 50021 0000 2 T1 TRANSFORMER 52005 000 1 U2 IC UPD75036 G 12 MASK ROM 54123 000 2 U1 IC MOG 3242...

Страница 31: ...nce performed every 18 months 1 Visually inspect the thermometer and probe for physical damage which might cause future product failure 2 Clean unit if necessary and replace batteries according to the...

Страница 32: ...7420 Carroll Road San Diego California 92121 800 854 2904 619 621 6600 FAX 619 621 6610 70017 000H...

Отзывы: