Diatek, A Welch Allyn Company
24
Model 600
temperature settings. The thermistor based instrument under test must be in Monitor mode and no
probe cover loaded. The probe is inserted into the small hole in the dry heat well of the 9600 and
allowed to settle for 2 minutes to the final temperature. The reading on the thermometer must be
within the range specified on the 9600. Refer to the M9600 Operation Manual for complete
instructions.
Note: All Diatek thermometers (thermistor and infrared ear thermometers can be tested in the
M9600.)
If you are having problems with the use of the M9600, refer to the Troubleshooting section in the
M9600 Operation Manual.
MODEL 600 TROUBLESHOOTING
GENERAL
SYMPTOM
PROBLEM
TREATMENT
No Display (Temp or Timer) or
low Battery Indication
Batteries Low or Incorrectly
Installed
Replace Batteries
No Temperature Display, Timer
OK
Probe Plug Bad or Not Seated
Probe Switch Not Activating
Reconnect or Replace Probe
Replace S1
Missing or Faded Display
Segments
Broken LCD Frame, Bad LCD,
LCD Connector Flex Strip,
Solder Bridge or U2
Locate and Repair
Broken Probe Indication
Defective Probe
Bad Q6 FET
Replace Probe
Replace Q6
Malfunction Indication
Unit Cannot Self Calibrate or
Microprocessor has a
Malfunction
Certain Broken Probe
Conditions
Clean Main PC Board
Replace Q3, Q4, Q5, C2, R3,
R4, R5, or U2
Replace Probe
Содержание Diatek 600
Страница 1: ...Diatek Model 600 Thermometer Technical Manual Thermo Menu Operation Manual...
Страница 13: ...Technical Manual Model 600 11 Figure 1 Block Diagram...
Страница 14: ...Diatek A Welch Allyn Company 12 Model 600 Figure 2 Schematic...
Страница 15: ...Technical Manual Model 600 13 Figure 3 Schematic earlier serial numbers...
Страница 16: ...Diatek A Welch Allyn Company 14 Model 600 Figure 4 Printed Circuit Board Component Layout...
Страница 17: ...Technical Manual Model 600 15 Figure 5 Final Assembly...
Страница 18: ...Diatek A Welch Allyn Company 16 Model 600...
Страница 19: ...Technical Manual Model 600 17 Figure 6 Operating Controls and Components...
Страница 32: ...7420 Carroll Road San Diego California 92121 800 854 2904 619 621 6600 FAX 619 621 6610 70017 000H...