Alarms
The monitor presents physiological alarms and technical alarms. Physiological alarms occur when
vital sign measurements fall outside of set alarm limits, but they occur only in the Monitor profile.
Technical alarms occur in all profiles.
Alarm types
Type
Priority
Color
Alarm audio tone
•
NIBP or SpO2 limit exceeded
•
Some technical alarms
High
Red
10-pulse tone
•
Pulse rate limit exceeded
•
Some technical alarms
Medium
Amber
3-pulse tone
•
Temperature limit exceeded
•
Some technical alarms
Low
Amber
2-pulse tone or 1-pulse tone
Alarm notification locations
WARNING
If you are relying on visual alarm notifications, maintain a clear line of
sight with the monitor and/or Nurse Call. If you are relying on audio alarm
notifications, ensure that you can hear audio alarms from where you are. Set the
volume as needed considering the environment and ambient noise levels.
Nurse Call
When the Nurse Call cable is connected and Nurse Call has been enabled, the monitor
immediately notifies the Nurse Call system when an alarm occurs. Nurse Call notification settings
are specified in the Advanced settings.
LED light bar
The light bar on the handle of the monitor illuminates as follows:
•
Flashing red for high priority alarms
•
Flashing amber for medium priority alarms
•
Constant amber for low priority alarms
49
Содержание Connex Vital Signs Monitor 6000 Series
Страница 1: ...Welch Allyn Vital Signs Monitor 6000 Series Directions for use ...
Страница 6: ...vi Contents Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 8: ...2 Introduction Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 14: ...8 Screen elements Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 22: ...16 Controls indicators and connectors Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 30: ...24 Setup Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 40: ...34 Navigation Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 44: ...38 Profiles Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 64: ...58 Alarms Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 92: ...86 Patient monitoring Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 98: ...92 Maintenance and service Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 106: ...100 Specifications Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 134: ...128 Troubleshooting Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 143: ......