measurements from the Home tab. Clinician
information clears from the Clinician tab and the
Device Status area.
Note
This setting does not take effect when
intervals are in progress.
Search by clinician ID
Enable the monitor to query the network for
clinician information based on ID. The monitor
initiates the search when the clinician enters or
scans the ID from the Clinician tab. Returned
clinician information populates the Device Status
area and fields on the Clinician tab.
Require password
Require clinicians to enter their password, in
addition to ID, on the Clinician tab. The monitor
uses the ID and password combination to query the
network for clinician information.
Require clinician ID match to save measurements Enable clinicians to save measurements only after
they have entered a clinician ID that matches an ID
in the network.
5. Do one of the following:
•
To continue in the Advanced Settings, touch another tab.
•
To exit the Advanced Settings and return to the Home tab, touch
Exit
.
Troubleshooting
Patient data management messages
Message
Possible cause
Suggested action
Unable to identify patient.
Touch Clear to delete all data.
The patient ID does not match an ID in the
patient list or network
Re-enter the patient ID.
To delete all unsaved data,
touch
Clear
.
Unable to identify clinician.
Touch Clear to delete all data.
The clinician ID does not match an ID in the
network
Re-enter the clinician ID.
To delete all unsaved data,
touch
Clear
.
Problems and solutions
The problems addressed in this table do not generate alarm or information messages on the
monitor.
Problem
Possible cause
Suggested action
•
Save button is disabled
•
Start intervals button is
disabled
No patient ID
Enter the patient ID.
Patient search is in progress (the Patient
frame displays a process indicator)
Wait until the search is
complete.
The patient ID does not match an ID in the
patient list or network
Re-enter the patient ID.
Addendum to the directions for use, software versions 1.0X–1.7X
3
Содержание Connex Integrated Wall System
Страница 1: ......
Страница 54: ...5 1 a b c d OK 50 1 0X 1 7X...
Страница 55: ...2 3 ID 4 ID 5 1 0X 1 7X 51...
Страница 56: ...52 1 0X 1 7X...
Страница 58: ...ID ID ID ID ID ID 5 ID 1 a b c d 2 3 ID 4 ID ID ID ID ID ID ID ID 54 1 0X 1 7X...
Страница 59: ...ID ID ID 5 ID ID ID ID ID ID ID ID ID ID ID ID ID ID ID ID ID 1 0X 1 7X 55...
Страница 60: ...56 1 0X 1 7X...
Страница 64: ...60 1 0X 1 7X...
Страница 68: ...Save Start intervals ID ID Patient ID ID ID ID ID Device Status Clinician ID ID ID ID ID 64 1 0X 1 7X...