Test the device
9
5. Connect the power supply testing set
MEDUMAT/MODULES WM 15440.
6. Connect CO
2
filter.
4.3 Enter device data
Enter the device number and the date of
manufacture in the test record.
4.4 Test leaktightness
4.4.1 Test oxygen duct
1. Apply a pressure of 6 bar to the inlet side of
both devices when they are switched off.
2. Shut off outlet pressure.
Requirement:
the drop in pressure must be
≤
200 mbar/min.
4.4.2 Test CO
2
measuring section
1. Push the CO
2
measuring and leaktightness
adapter onto the intake opening for respiratory
gas CO
2
measurement
ensure that the CO
2
measuring and
leaktightness adapter is correctly positioned
(the pin must be in the position shown in
relation to the device).
2. Put the T-tube with syringe from set
WM 15349/WM 15359 onto the CO
2
measuring and leaktightness adapter.
3. Detach the test pressure gage from the T-tube.
4. Connect the volume flow measuring device at
the same point.
5. Push the CO
2
measuring and leaktightness
adapter against the intake opening for
respiratory gas CO
2
measurement
device to ensure leaktightness.
6. Apply a pressure of 60 mbar to the intake
opening for respiratory gas CO
2
measurement
Requirement:
the drop in pressure must be
≤
2 mbar/min.
Pin
CO
2
measuring and
leaktightness adapter
T-tube with syringe