Test the device
13
4.12 Test CO
2
measurement
4.12.1 Preparation
1. Plug water filter
respiratory gas CO
2
measurement
2. Connect CO
2
absorber WM 22172 to water
3. Press the CO
2
absorber at least 20 times to
ensure that CO
2
-free air is in the system.
4.12.2 Method
Important!
Use only the following components:
• the KAL Check gas (WM 97061) supplied by
WEINMANN
• a suitable pressure reducer (WM 97062)
Observe an ambient temperature of approx. 20 ˚C to guarantee adequate precision in this test. It is essential
to follow the instructions for use and the warning information enclosed with the KAL CHECK gas.
1. Switch on the device.
Leave MODUL CapnoVol to run for approx.
15 minutes in normal mode.
2. Press the “Menu” key. The main menu
appears.
3. Use the arrow keys to select the menu item
“CO2 measurement”.
4. Press the
↵
key to reach the submenu for CO
2
measurement.
5. Use the
‹
key to select the “Calibration check”
submenu.
6. Press the
↵
key to reach the submenu.
7. Disconnect CO
2
BiCheck flow sensor
Содержание Emergency MODUL CapnoVol