Repair information and repair instructions
27
8. Hold bottom part of the housing
and the
circuit boards in one hand, turn bottom part of
the housing
Caution!
When fitting the power supply connection, the
Capno connection adapter and the connection
for the BiCheck flow sensor, ensure that the
connections are correctly positioned in the
corresponding ducts in the housing (groove
and spring), otherwise the connections or the
bottom part of the housing may be damaged.
9. Push power supply connection
corresponding bore of bottom of the housing
and screw it up with a hexagon nut
10. Push Capno connection adapter
corresponding bore of bottom part of the
housing
described in Section “7.4 Replace the Capno
connecting adapter” on page 24.
11. Push BiCheck flow sensor connection
into the
corresponding bore in the bottom of the
housing and screw it up with hexagon nut
12. Put battery
in the bracket in the bottom of
the housing and connect the connector of
battery
X1
on the circuit
board.
13. Close the device (see “7.3 Close the device”
14. Test the device (see “4. Test the device” on
Note
Return the faulty circuit boards to
WEINMANN.
X1
Содержание Emergency MODUL CapnoVol
Страница 43: ......