40
Technical data
EN
11. Technical data
11.1 Specifications
CPAP 20e
CPAP 20e with CLICK 2
Product class as per
93/42/EEC
IIa
Dimensions WxHxD in
cm
21 x 9.8 x 18.5
21 x 14 x 31.3
Weight
approx. 1.25 kg
approx. 1.5 kg (without water)
Temperature range
– operation
– storage
+5 °C to +35 °C
–20 °C to +70 °C
If the unit runs at +40 °C , the air output can heat up to 42 °C.
Adm. humidity for
operation and storage
≤
95 % rF (no condensation)
Air pressure range
600 - 1100 hPa (automatic altitude compensation)
(allows for operation in an altitude of up to approximately 4000 meters)
Diameter of breathing
hose connection (mask)
in mm
19.5 (suitable for 22 mm standard cone)
Electrical connection
115-230 V AC +10/-15 %, 50–60 Hz with WM 24480
power supply or 12-24 V DC -15/+25 % with WM 24469 DC adapter
Power consumption
during
– operation
– standby
230 V
115 V
24 V
12 V
0.1 A
0.2 A
0.8 A
1.6 A
0.02 A
0.03 A
0.2 A
0.24 A
230 V
115 V
24 V
12 V
0.14 A
0.28 A
1.3 A
4.0 A
0.02 A
0.03 A
0.2 A
0.24 A
Classification as per
prEN 60601-1:2004
– type of protection
from electric shock
– level of protection
from electric shock
– protection against
damaging ingress of
water
– operating mode
Protection class II
Type B
IPX1
Constant operation