Functional check
33
4. When the functional check has finished, the
message ”Function check OK” appears and
BiLevel ST 22 reverts to the standard display.
If a plausibility error is found during the functional
check, the message “Flow sensor implausible!”,
“Sensors implausible!” or “Implausible pressure
measurement!” appears.
Close the window using the menu key
, turn the
BiLevel ST 22 off and disconnect the power for
5 minutes. Check whether the drying adapter has
been fitted correctly. Repeat the functional check.
If the message “Flow sensor implausible!”, “Sensors implausible!” or “Implausible
pressure measurement!” appears again, please contact Weinmann or your
dealer immediately to have your unit repaired.
Alarms
BiLevel ST 22 carries out a
self-test on the sensors when the on/off switch is
pressed. The alarm feature is also tested. If an error occurs during the self-test, an
error message appears on the main screen (see also "7. Troubleshooting" on
page 34).
1. Check the buzzer and the status displays:
Each time you turn the BiLevel ST 22 on, check that you can hear the buzzer and
that the status display lights up.
2. Check the power failure alarm:
Start BiLevel ST 22. Now pull the power supply cable out of the socket; the
display goes out and the buzzer sounds.
Note
BiLevel ST 22 must have been connected to the mains for at least 5 minutes
before carrying out the test.
Turn the unit off after the functional check has been completed by pressing the on/
off
switch.