Setting up the unit
Bacteria filter
If BiLevel ST 22 is to be used by several patients (e.g.
in a hospital), the WM 24148 bacteria filter must be
used to protect against infection. It is connected
between the hose system and the BiLevel ST 22 or
VENTI
click
. Also refer to the Operating Instructions
supplied with the bacteria filter.
Caution!
The bacteria filter provides additional resistance to the air flow: This can bring
about a change in how the trigger responds. The unit parameters must therefore
be adjusted if a bacteria filter is added later.
VENTI
power
The VENTI
power
is available for mains-independent
operation.
VENTI
power
can be connected to the therapy unit as
an alternative to the mains supply.
Also refer to the Operating Instructions for the
VENTI
power
.
Note
We advise against using the humidifier at the same time as the VENTI
power
, as
this considerably reduces the duration of the power supply from the VENTI
power
.