15
example
two decimal point
[C02
2]
4
press
press
press
or
[C03
]
[C03
1]
[C03
5]
Set graduation
option
1/2/5/10/20/50
Select required graduation
example
graduation 5
[C03
5]
5
press
press
press
or
/
[C04
]
[0100.00]
[0100.00]
Max capacity
example
max weighing 100kg:
[0100.00]
6
press
press
press
press
[C05
]
[C05
0]
[C05
1]
[CAL
9]
[0000.00]
Zero calibration
Option
0=no need zero calibration
1=need zero calibration
calibration zero please choose 1 and
ensure scale is empty and “stable”
light is on
Ensure zero calibration, countdown.
Till
show[0.00](example
for
two
decimal point)
7
press
press
press
or
press
press
or
press
[C06
]
[C06
0]
[C06
1]
[SPAN ]
[0100.00]
[0080.00]
[CAL
9]
……
[0080.00]
[CAL End]
calibration
option:
0=No need calibration
1= need calibration
Load
weights
on
scales
according
to
max.
capacity.
Suggest
close
to
the
max
capacity,
at least 10% of max.
capacity.
For example: the weights is
80kg
As bellows:
Input the 0080.00, count down ,