Connection
Î
Grip the connector by its sliding sleeve (Pos. 1). Make sure that the clamping jaws
are open and protrude from the sliding sleeve (Pos. 1).
Î
Be careful to place the connector straight onto the test piece. Doing so will prevent
any damage to the front seal caused by any sharp edges on the test piece and en-
sure a correct connection.
Î
Push the connector at the sliding sleeve (Pos. 1) forward
onto the test piece (
Figure 1
). The connector must engage
with a clearly audible click. The retaining ring must be visible
(
Figure 2
) and the outer edges of the external diameter of the
clamping jaws must no longer be visible (
Figure 3
and
Figure 4
).
Attention:
If the connector does not click into place, the
outer edges of the external diameter of the clamping jaws
are visible or the circlip is not visible, then the connector is
not correctly connected. The connector may spring out of
place and medium may escape.
Î
Pull gently on the lower part of the connector (Pos. 2)
(red arrow
Figure 5
) in the direction opposite that of the
connection to make sure that the connector is properly
connected.
- The connector is now connected pressure-tight to the test
piece.
Î
To start the testing procedure, apply pressure to media
inlet ‘B1’.
Figure 1
Figure 4
Figure 3
Figure 2
visible
Figure 5
pull backwards
MD-10060-L51-R1.1.0-03