SLS Baseline 9000 FID Total Gas Detector Manual
7)
After the zero value has been set
(or skipped
), the machine will automatically move to the span
calibration. Connect the 10% methane calibration gas to the span port at 20 psi.
Figure 30: Span Connection
8)
Use the left button to update the span value until the reading has settled
, likely
fluctuating
above and below the span value. When the reading has settled,
press the green enter key
to
save the calibration. This process may
be skipped by pressing the red escape key if no span
calibration is required.
Figure 31: Span Calibration