Page
14
of
36
SLS Baseline 9000 FID Total Gas Detector Manual
8)
Measure the flow out of the detector vent on the back panel and compare the measured value
with the target flow shown on the display. If necessary, use the
← key and the numeric keypad
to change the target pressure counts. Use the
→ key to update the new value. Wait for the
pressure counts to stabilize, at least 1 minute, and re-measure the flow at the detector vent.
Repeat until the measured flow is equal to the target flow ±2 cc/min.
Figure 12: Fuel Flow Value
Figure 13: Fuel Flow Reading
9)
When the measured flow equals the target flow, press “Enter” to save the span point.
10)
If you make a mistake, hit Escape to exit ou
t without saving the changes.