SLS Baseline 9000 FID Total Gas Detector Manual
7)
Wait until the current pressure counts are equal to or close to the target pressure counts,
minimum 2 minutes.
Figure 16: Air Flow Press Cnts
8)
Measure the flow out of the detector vent on the back panel and compare
the measured value
with the target flow shown on the display. If necessary, use the ← key and the numeric keypad
to change the target pressure counts. Use the → key to update the new value. Wait for the
pressure counts to stabilize, at least 1 minute, and re-measure the flow at the detector vent.
Repeat until the measured flow is equal to the target flow ±2 cc/min.
Figure 17: Air Flow Value