background image

9

Care Equipment Excellence

www.wealdenrehab.com 

     wealdenrehab 

01634 813388 
[email protected]

5 Maintenance & Cleaning

Warning: 

Before carrying out any maintenance, remove the battery.

Routine maintenance can be carried out by users and/or assistants who are familiar with these instructions.
A LOLER Test Certificate is required every 6 months and should be carried out by a qualified technician.
The inspection must include a visual and functional check on the following elements:
• Structure, anchors, components, mechanisms, controls, safety systems;
• Supporting devices for the individual (slings, spreader bar, seat, mobile base, etc.)

Check the safety of the equipment after any “rough treatment” (e.g..: falls, bumps, etc.) and whenever in doubt.
Use a cloth and a mild detergent to clean the painted parts.
To prevent oxidation of chromed parts, every 2 weeks clean them and all the parts subject to friction (e.g..: actuator sleeve,
arm-seat anchoring unit, etc.) with a stainless steel cleaner. Batteries are sealed and therefore maintenance free, they only need to 
be recharged.

Warning: 

The equipment is not protected against water spray.

Troubleshooting

Problem

Cause

Solution

The equipment won’t work

Emergency button is ON

Disable

Connectors disconnected or improperly 
plugged in

Check connections

Flat batteries

Recharge

Burnt battery protection fuse

Contact service centre

Damaged control box

Contact service centre

Batteries do not recharge

Battery charger is not powered or is  
improperly connected

Check connections
Check mains voltage

Damaged battery charger

Replace

Spent batteries

Replace

Equipment stuck in raised position

Damaged control box – remote control  
or actuator

Use emergency lowering
Contact service centre

The equipment keeps moving although 
the buttons are not pressed

Damaged control box – remote control

Use emergency stop button
Contact service centre

Control box acoustic signal triggered

Batteries at 50% of charge

Recharge as soon as possible

Technical assistance & spareparts

Wealden Rehab offers technical assistance by phone. 

01634 813388

. Outside business hours you can email us at 

[email protected]

. Here follows a reference list of the spare parts that are most subject to wear and tear.

Code

Description

Quantity

1

N4954

CBJ2 IP65 Grey Control Box

1

2

H2061

BAJ2 IP41 Grey Battery Pack

1

3

H2062

Linak Battery Charger

1

4

N4955

HB71000 IP66 Grey Remote Control

1

5

N4953

LA34 Grey Actuator

1

Содержание N3242-1 Heron Lift E

Страница 1: ...ehab com Heron Pool Lift Operation Maintenance Manual N3242 1 Heron Lift E with spreader bar N3242 3 Heron Lift Max E with spreader bar N3242 0 Heron Lift E with integral seat N3242 2 Heron Lift Max E with integral seat N5429 Mobile Base For Liftpoolseat N2463 Pool Socket N2465 Pool Plate ...

Страница 2: ...ental conditions 3 Safety Page 4 General warning Proper and improper use Emergency lowering and stop functions Setting up for use Installing the anchor systems Main components and dimensions cm Use 4 Battery Charging Page 4 5 Maintenance Cleaning Page 9 Troubleshooting Technical assistance and spareparts 6 Lifting Supports Page 10 Lifting supports 7 Warranty Specification Conformity Page 10 Warran...

Страница 3: ... failure to comply with these instructions Target readership This manual is intended for users therapist attendant it must always be kept with the equipment and stored in a safe place for future reference 2 Equipment Description Intended use operation Electric hoist for easy transfer of disabled users into a swimming pool Painted stainless steel Available versions HERON with a sling bar and HERON ...

Страница 4: ... if no lifeguard is at the pool Emergency lowering stop functions The machine can be stopped by depressing the lift lower buttons pressing the emergency button Fig 3 to disable the button and put the equipment back into operation rotate the button in the arrow direction Warning Press the emergency button when you need to stop the equipment immediately The emergency lowering can be enabled by press...

Страница 5: ...tem must be capable of withstanding the load applied Pool Socket Determine the proper position for installing the hoist see fig 2 Bore a hole 90mm in diameter and 240mm deep Put a shim on the bottom of the hole so that the socket and cap are level with the floor Remove the socket and fill 1 4 of the hole with fluid quick cement Insert the socket and press it until the cap is flush with the floor M...

Страница 6: ...he first time 7 Connect the remote control and the actuator connectors to the control box Fig 3 Make sure that the pins are fully inserted otherwise the hoist may not work Use Heron with spreader bar 1 Apply the sling to the patient refer to special instructions 2 Press the Lower button on the remote control until the desired height is achieved and connect to the sling 3 Raise the patient by press...

Страница 7: ...ase item N5429 for Heron Seat The mobile base is an accessory to be applied to the seat of the HERON to transfer the patient from the changing room to the hoist and vice versa The mobile base is supplied with the seat To attach the seat to the hoist do as follows Fig 4 1 Transfer the patient onto the seat mobile base with the wheel brakes engaged fold back the safety armrests 2 Unfold the footrest...

Страница 8: ...d battery upwards 2 Connect the power supply unit to the mains with a voltage of 220V 50Hz and insert the charging pin into the relevant socket located on the battery 3 When charging is completed disconnect the battery charger and reinsert the battery into its seat control box until it clicks into place Battery life can be extended if you avoid discharging batteries completely We recommend you to ...

Страница 9: ...water spray Troubleshooting Problem Cause Solution The equipment won t work Emergency button is ON Disable Connectors disconnected or improperly plugged in Check connections Flat batteries Recharge Burnt battery protection fuse Contact service centre Damaged control box Contact service centre Batteries do not recharge Battery charger is not powered or is improperly connected Check connections Chec...

Страница 10: ...odge a complaint please contact our Technical Assistance Manager Technical sheet N3242 1 N3242 3 N3242 0 N3242 2 N2463 N2465 10 4 14255 N5429 Item name Electric Liftpool Electric Liftpool seat without mobile base Socket Plate Legrest Mobile base without seat Dimensions cm See figures Weight 35 38max 55 58max 5 22 5 10 Max capacity 135 135 Use Equipment for temporary use 10 max 6min hour Electric p...

Страница 11: ...ldenrehab 01634 813388 hello wealdenrehab com 8 Inspection record sheet Equipment Manufacturer CHINESPORT S p A 33100 Udine Italy Serial number Date of first installation Date of inspection Operations carried out Technician s name Signature Next due date ...