![Waygate Technologies Krautkramer USM Go Скачать руководство пользователя страница 91](http://html.mh-extra.com/html/waygate-technologies/krautkramer-usm-go/krautkramer-usm-go_operating-manual_878369091.webp)
Setting the pulser (function group PULSER)
5 Operation
USM Go
Rev. 10 (06/2021)
5-15
ATTENTION
Use the data sheet for your probe to check
which maximum voltage is allowed to be ap-
plied.
Note
The pulser voltage and the pulse width can
be automatically limited, depending on the
pulse repetition frequency (see Section
MODE (pulse repetition frequency)
page 5-17) setting. This function helps to
avoid heat accumulations in the pulser elec-
tronics.
– Select the function
VOLTAGE
.
– Press the function keys to choose the required pulser
voltage.
ENERGY
You can use the function
ENERGY
to set the penetra-
tion or sound energy.
The following settings are possible:
●
HIGH
– high energy
●
LOW
– low energy
The setting
HIGH
is recommended for all tests in which
maximum sensitivity is important, e.g. for the detection
of small flaws. Choose the setting
LOW
for broadband
probes or if narrow echoes are required (better lateral
resolution).
– Select the function
ENERGY
.
– Press the function keys to choose the required set-
ting.
Содержание Krautkramer USM Go
Страница 1: ...Krautkrämer USM Go USM Go Operating Manual Id No 49 155 Rev 10 06 2021 ...
Страница 5: ...USM Go Rev 10 06 2021 0 5 Second operating level ...
Страница 6: ...0 6 Rev 10 06 2021 USM Go Second operating level continued ...
Страница 10: ...0 10 Rev 10 06 2021 USM Go ...
Страница 20: ...0 10 Rev 10 06 2021 USM Go ...
Страница 21: ...USM Go Rev 10 06 2021 1 1 Introduction 1 ...
Страница 35: ...USM Go Rev 10 06 2021 2 1 Standard package and accessories 2 ...
Страница 41: ...USM Go Rev 10 06 2021 3 1 Initial start up 3 ...
Страница 53: ...USM Go Rev 10 06 2021 4 1 Principles of operation 4 ...
Страница 76: ...4 Principles of operation Saving the settings 4 24 Rev 10 06 2021 USM Go ...
Страница 77: ...USM Go Rev 10 06 2021 5 1 Operation 5 ...
Страница 210: ...5 Operation Evaluation according to the DGS method 5 134 Rev 10 06 2021 USM Go ...
Страница 211: ...USM Go Rev 10 06 2021 6 1 Documentation 6 ...
Страница 240: ...6 30 Rev 10 06 2021 USM Go ...
Страница 241: ...USM Go Rev 10 06 2021 7 1 Maintenance and care 7 ...
Страница 247: ...USM Go Rev 10 06 2021 8 1 Interfaces and Peripherals 8 ...
Страница 251: ...USM Go Rev 10 06 2021 9 1 Appendix 9 ...
Страница 266: ...9 16 Rev 10 06 2021 USM Go ...
Страница 267: ...USM Go Rev 10 06 2021 10 1 Specifications 10 ...
Страница 278: ...10 12 Rev 10 06 2021 USM Go ...
Страница 279: ...USM Go Rev 10 06 2021 11 1 Index 11 ...