background image

14

OPERATING INSTRUCTIONS - LDTC LAB

INSTALL THE NATIONAL INSTRUMENTS 
RUNTIME ENGINE 2017

Prior to installing Wavelength's Virtual Instrument application, 

the National Instruments (NI) Runtime Engine 2017 must be 

installed on the remote computer.

When  installing  the  Virtual  Instrument  .EXE  file,  if  the 

Runtime Engine is not already installed on the computer, a 

message with a link to the NI website download page opens. 

The Runtime Engine can take several minutes to install and 

requires restarting the computer.

INSTALL LDTC LAB VIRTUAL INSTRUMENT 
APPLICATION

1.  On the 

Software Downloads page

, click the LabVIEW 

Interface for LDTC LAB Series Instrument and download 

it to the computer.

2. 

Save the file to the computer, and open it.

3.  Follow the prompts to install the application.

CONNECT THE LDTC LAB TO THE REMOTE 
COMPUTER

Connect a USB cable from the instrument to the remote 

computer, turn on the rear panel switch and open the Virtual 

Instrument application.
Choose the instrument to control using the 

Choose Instrument

 

dropdown  menu  on  the  right  hand  side.  The  list  is  filtered, 

and will only show LDTC LAB instruments connected to the 

computer.

Figure 15.  Choose the LDTC LAB to control.

The other functionality of the executable is disabled until a 

proper connection is made with an instrument. 

Start:

 Use this button to initialize the program 

after selecting the instrument.

Once a connection is made, the program will initialize all the 

parameters to match the local values on the instrument.

ADDITIONAL CONTROLS

Stop:

 Use this button to gracefully stop the 

execution of the program. 

Run:

 This button will only be available if the program has 

been stopped previously. Use it to re-start the execution 

of the program. This is located in the upper left corner 

of the window.

Running: 

This icon will only appear when the program 

is currently running after pressing the 

Start

 button. This 

will replace the 

Run

 button in the upper left corner.

USING THE GUI

The graphical user interface (GUI) for the LDTC LAB allows 

the user to set and monitor the status of the instrument 

remotely. 

Figure 16

 gives an example of the GUI.

Numerical monitor-only values have a gray background, and 

are uneditable fields (for example, see Actual Temperature). 

Boolean monitors are dark for FALSE and lit for TRUE (see In 

Tolerance) and are also uneditable.
Button controls set a boolean value, and also indicate the 

current value of their state (for example, see TC ENABLE).
Controls have a white background, and are editable either 

through  typing  the  desired  value  directly  in  the  field,  or  by 

choosing the desired setting from a dropdown box (see 

Temperature Setpoint and Temperature Units). When a 

value is set using a control, the instrument is immediately 

queried,  and  the  field  is  updated  with  the  current  status  of 

the instrument. In this way, the GUI maintains synchronization 

with the instrument.

Figure 16.  Various controls and monitors on the GUI.

TC ENABLE: TC current is enabled.

Actual  Temperature:  Read  only  field  -  the  temperature  of 

the currently used sensor. Actual Temperature shown is 

18.0001ºC.

Temperature Units: Dropdown box - choice is to display 

temperature in degrees Celsius.

TEC I (A) & TEC V (V): Read only fields - TEC current and 

voltage. The TEC Current shown is 0.6239 A, and the TEC 

Voltage shown is 0.8573 V.

Temperature Setpoint: Editable field - 18.0000ºC has been 

chosen for the setpoint.

In Tolerance: Indicator is lit, so Actual Temperature is within 

tolerance settings of setpoint.

INSTRUMENT 

SERIAL NUMBER

Содержание LAB Series

Страница 1: ...bit RAM 256 MB minimum Screen Resolution 1024 x 768 pixels Disk Space 5 GB available hard disk space 32 bit 20 GB DirectX 9 graphics device with WDDM 1 0 or higher driver 64 bit Installation Prerequis...

Страница 2: ...s Virtual Instrument application the National Instrument NI Runtime Engine 2015 must be installed on the remote computer When installing the Virtual Instrument EXE file if the Runtime Engine is not a...

Страница 3: ...below the value specified Cable Resistance In the Cable Resistance control type the resistance of the cable in Ohms between the QCL LAB instrument and the load The QCL voltage displayed will be VDISPL...

Страница 4: ...can with decreasing current Step Duration sec In the Step Duration control type the time delay in seconds between each step in the VI scan The step size and duration determines the length of the scan...

Страница 5: ...logged Logged Data File Header In the Logged Data File Header control type the heading that appears at the top of a new data file If appending data the file header is not rewritten ADDITIONAL SETUP TA...

Страница 6: ...onal Instrument NI Runtime Engine 2015 must be installed on the remote computer When installing the Virtual Instrument EXE file if the Runtime Engine is not already installed on the computer a message...

Страница 7: ...rent Limit Power The Power button functions the same as the front panel power switch on the instrument Enable The Enable button enables current to the load and functions the same as the Enable button...

Страница 8: ...ain lit until the scan is complete Clicking the button again after starting IntelliTune will not stop the cycle The IntelliTune Abort button must be used to stop a cycle in progress If IntelliTune doe...

Страница 9: ...ntrol will configure the bias current for the auxiliary sensor Choose Auto to allow the TC LAB auto range function to choose an appropriate bias current or select a suitable range that will produce a...

Страница 10: ...istance sensors T1 R1 T2 R2 T3 R3 need to be populated by the temperatures resistances for the first second and third temperature resistance pairs respectively For Temperature Voltage sensors T1 V1 T2...

Страница 11: ...remote temperature scan the temperature iterations are done within the VI and as such no data will be logged to a flash drive if inserted The data will be saved to the appropriate file directory as sp...

Страница 12: ...efore data can be logged Logged Data File Header In the Logged Data File Header control type the heading that appears at the top of a new data file If appending data the file header is not rewritten I...

Страница 13: ...output The signal will go to the level indicated when the Actual Temperature exceeds the temperature limits Options are High 5V and Low 0V The temperature limits are configured on the Settings Monito...

Страница 14: ...has been stopped previously Use it to re start the execution of the program This is located in the upper left corner of the window Running This icon will only appear when the program is currently runn...

Страница 15: ...nded by Temp Limit Min and Temp Limit Max InTolerance Thisbooleanindicatorislitwhenthetemperature is in tolerance as described by the Temp Tolerance and Temp Time Tolerance fields INTELLITUNE PARAMETE...

Страница 16: ...temperature limit here in units chosen by the Temperature Units dropdown Available range is 99 to 250 C This value must be less than Temperature Setpoint and Temp Limit Max Temp Limit Max Enter the h...

Страница 17: ...ed by the command LASer AMPS returning 1 true If the indicator is not lit the units are in mA This will affect Laser Setpoint Laser Current Laser Current Limit Laser Current Tolerance Start Current St...

Страница 18: ...it This indicator is lit if the laser driver is operating in a current limit condition Time Delay Prior to Enabling LD Current msec Sets the delay time in milliseconds between when the output on comma...

Страница 19: ...oefficients Callendar Van Dusen parameters or Slope Offset parameters that can be used to characterize the sensor Save Configured Sensor Press the SAVE button when the calibration parameters the Senso...

Страница 20: ...profile has been selected press RECALL to load the profile The application will write the values to the instrument pause and re initialize to ensure synchronization Save to Profile Choose to which pro...

Страница 21: ...Data Temperature Scan Plot This plot shows the collected data Data points are Actual Temperature white circular points and Setpoint Temperature blue square points TEMPERATURE SCAN Start Temperature I...

Страница 22: ...t when the Begin V I Scan button has been pressed but the instrument is waiting for Laser Current to reach stability at the start current The scan will abort if the current does not come in to toleran...

Страница 23: ...Temperature red Low Temperature Limit green High Temperature Limit blue Temperature is plotted in the units currently active on the instrument LASER This plot shows various laser control parameters a...

Страница 24: ...0 A 100 A 1 mA or 10 mA indicating the bias current value IntelliTune Mode One of the mode indicators will be lit indicating the current IntelliTune mode setting LD STATUS Laser Type Control Mode Ther...

Страница 25: ...any other custom application without the express written permission of Wavelength Electronics Inc The QCL TC LDTC LAB LabVIEW Interface applications and their components and modules may not be redistr...

Страница 26: ...ry to change the temperature in the TC LAB VI but the VI resets the value to a different temperature A Temperature Scan is in progress Either wait for the temperature scan to finish or abort the tempe...

Страница 27: ...Buyer End User or Third Party Reseller are expressly prohibited from reverse engineering decompiling or disassembling this product NOTICE The information contained in this document is subject to chang...

Отзывы: