2.0 Overview
The CLX-Ex Online Chlorine Monitor allows for the reading of chlorine
levels of process water on-line. The CLX-Ex has been designed
to meet the design criteria specified by Standard Methods for the
Examination of Water and Wastewater (21th Edition) Method 4500-Cl
G. DPD Colorimetric Method. The CLX-Ex uses a 515nm LED as the
measurement light source.
This instrument has been designed to operate in explosive
atmospheres and meets the IECEx rating as stated in the
specifications. The instrument uses an approved purge/pressurization
controller called the Air Sentinel. Please refer to the companion manual
Cat. No. 28749 for all information on:
• Operation of the Air Sentinel
• Electrical power connections
• Purge Air Connection
*Viton
®
is a trademark of The Chemours Company FC, LLC
Noryl
®
is a registered trademark of SABIC Global Technologies B.V.
2.1 Specifications
Measurement Range
0.00 – 10.00 mg/L (PPM)
Accuracy
±5% of reading or ±0.03 mg/L (PPM) whichever is greater for range of 0-6.0 mg/L(PPM) ±10% of reading from 6.01-10.00 mg/L (PPM)
Resolution
0.01 mg/L (PPM)
Cycle Time
Adjustable; 60 seconds to 10 minutes (600 seconds)
Display
Multi-Line Liquid Crystal Backlit Display
Alarms
One NO relay contact 120-240VAC 2A
Analog Output
Powered 4-20 mA, 600 Ω drive, isolated
Communications Port
Bi-directional RS-485 with Modbus
Water Pressure
Integral pressure regulator 0.34 bar (5.0 PSI) to 10.3 bar (150 PSI.)
Flow Rate to Waste
200 – 400 ml/min.
Operating Temperature
0°C – 55°C (32°F – 131°F)
Wetted Materials
PVC, Borosilicate Glass, Reslyn (FFKM), Viton
®
(FKM), Polypropylene, 316 Stainless Steel, Acetal, Noryl
®
, Silicone
Sample Temperature Range
0°C – 55°C (32°F – 131°F)
Power Supply
100 to 240VAC, 47-63 Hz, 250VA
Insulation Rating
Double Insulated, Pollution Degree 2, Overvoltage Category II
Environmental Conditions
Altitude up to 2000 meters
Up to 95% RH (non-condensing)
IECEx Hazloc Rating
Ex pxb IIC T4 Gb 0°C ≤ Ta ≤ 55°C
Compressed Air
Water and oil free, -40°F (-40°C) Dew Point, Particles <5u, ISA Grade Hydrocarbon Free. Full time clean dry air at 5.5 - 7 bar
(80-101.5 PSI) @35 SLPM (1.2 SCFM) @ 20°C (68°F) Max
Regulatory Compliance and Certifications
ETL listed to UL 61010-1 Issued May 11 2012 3rd Ed and ETL certified to CSA 22.2 No. 61010.1 issued May 11 2012
EN61326:2006
Shipping Weight
27.2 kg (60 lbs.) Reagents are Shipped Separately
Shipping Dimensions
93 cm X 62 cm X 32 cm (36½" X 24½" X 12½")
In addition this instrument has vortex cooling. This instrument will
automatically control a connected compressed air line to lower the
temperature of one of the reagents, allowing the reagent an operating
life of up to 3 months.