IS-HF-M100+-29293  Rev 4

 

4

 

Introduction to the Unit

Overview

The M100+ is a process turbidimeter that allows for the laboratory  

measurement of the turbidity of sample water. The white light M100+ 

has been designed to meet the design criteria specified by the US EPA 

180.1 on turbidity measurement. The infrared M100+ was designed 

to meet the criteria specified in ISO 7027 and DIN 27027 for the 

measurement of the turbidity of a sample. Both models have long life 

lamps.
The M100+ instruments are equipped with a USB slot in order to 

download logged data and calibration information. The USB slot is also 

used to update operating software.
The M100+ is equipped with a color touch screen for easy and efficient 

operation.

Specifications

M100+ White Light 

Model 28060

Conforms to specifications set forth in EPA method 180.1 

(Nephelometric Method)

§

 This model uses a tungsten 

filament lamp.

M100+ Infrared 

Model 28061

Conforms to specifications set forth in ISO 7027: Water 

Quality - Determination of Turbidity. This model uses an 

IR LED lamp.

Measurement Range

0-4000 NTU

Accuracy

† ‡

±

2% of reading or 

±

0.02 NTU below 40 NTU, whichever 

is greater

±

5% of reading above 40 NTU

Repeatability

±

1% of reading or 

±

0.02 NTU, whichever is greater

Resolution

Menu settable up to 0.0000 over the entire range

USB

Printer or download logged data & calibration data & 

upload new firmware via thumb drive (USB flash stick).

Power Supply

UL, CSA & CE approved 12V DC, Wall Mount

Miscellaneous Specifications

1. USB port for logged data download and software 

updates.

2. Built-in diagnostics

3. Three year battery backup with no external power

Operating Temperature Range

1°C to  40°C (34°F to 104°F)

Sample Temperature Range

0°C to 40°C (32°F to 104°F)

Dimensions

237mm W x 254mm L x 121mm H 

(10.75”W x 10” L x 4.75” H)

Shipping Weight

2.5 kg (5.5 lbs)

Certifications

LC mark tested to UL and CSA. Conforms to CE.

Warranty

1 year from date of shipment

§

 The specifications found in EPA method 180.1 are essentially the same as the specifications set out 

in method 2130B of the Standard Methods of the Examination of Water and Wastewater 22nd Edition 

and the specifications set out in ASTM Standard Method D1889-94. The M100+ meets or exceeds 

the specifications set forth in these methods.

 Instrumental accuracy measured under controlled laboratory conditions at 25°C (77°F).

 Both the accuracy and repeatability specifications for the M100+ are valid only for measurement  

of static (non-flowing) samples.

Touch Screen

Standard Inserted in Optics

USB Connector

Unpacking and Inspection of the Instrument

Remove all items from packing carton and carefully inspect to ensure 

that no visible damage has occurred during shipment and that all items 

listed below have been received. If the items received do not match 

the order, please contact your local distributer or the HF scientific 

Customer Service department.

•  M100+ Laboratory Turbidimeter
•  Accessory Kit for M100+ (2 empty sample cuvettes with caps) 
•  Full Range Calibration Kit High Range A, High Range B,  

1000 NTU, 10 NTU & 0.02 NTU

•  Quick Start Guide
•  Flash Drive - Contains full owner’s and user’s manual
•  Wall Mount Power Supply with Ferrite

A ferrite is included with the power supply. This should be 

looped twice through the cord near the side that connects to 

the instrument to eliminate EMI interference.

NOTICE

Site Selection

The follow should be taken into consideration when selection a site for 

the M100+ instrument.

•  Must be in a suitable, dry location with adequate ventilation.
•  Should not be located where chemicals, such as chlorine, are 

located. 

•  Must be operated and stored between 1 and 40°C. 
•  Should be used on a stable table top.
•  Table top location should be less than 90 cm (3 feet) from a 

power source.

Power Requirements

The M100+ requires 12VDC for proper operation. The provided power 

supply is rated at 1A. 
The instrument is supplied with a wall mount power supply intended 

for oepration with 100-240 VAC, 47-63 Hz power source. 

Before 

installing, verify that the line voltage falls within these 

specifications.

The wall mount power supply connects to the instrument’s rear panel.
An optional Euro wall mount power supply is available.

Setup

Before using your M100+, you will need to familiarize yourself with the 

instrument’s operating modes and functions and perform some setup, 

configuration and calibration.
The M100+ allows for the measurement of the turbidity of water. The 

turbidity of the water is usually reported in Nephelometric Turbidity 

Units (NTU) but may be reported in Formazin Nephelometric Units 

(FNU). This is user selectable.
Readings up to 4400 are possible but readings above 4000 are outside 

of the rated specifications. Readings above 4400 NTU will cause the 

display to flash, indicating an over-range condition.
During normal operation the instrument will be at the HOME screen. 

This is the screen where measurements can be viewed.
Please note, some features depicted in this manual may change with 

future firmware updates.

Содержание M100+ 28060

Страница 1: ...en Calibration 6 Set Date Time 6 Delete Samples 6 Data Resolution 6 Offset 7 Device ID 7 Calibration Reminder 7 Calibration Download 7 Firmware Update 7 Reset Defaults 7 About Device 7 Calibrating the M100 8 Calibration Standards 8 Care of ProCal Standards 8 Indexing Calibration Cuvettes 8 Calibration Procedures 8 Calibration Procedures continued 9 Calibrating using ProCal Standards 9 Calibrating ...

Страница 2: ...hould not permit anyone to touch this equipment unless they are over 18 years of age are adequately trained and supervised and have read and understood this manual Owners should also ensure that no unauthorized personnel comes in contact with this equipment If this equipment or any of its parts becomes damaged or needs repair stop using the equipment and contact an experienced service individual i...

Страница 3: ...lt in death or serious injury This symbol identifies hazards which if not avoided could result in death or serious injury This symbol identifies hazards which if not avoided could result in minor or moderate injury This symbol identifies practices actions or failure to act which could result in property damage or damage to the equipment This pictorial alerts you to the need to read the manual This...

Страница 4: ...ons for the M100 are valid only for measurement of static non flowing samples Touch Screen Standard Inserted in Optics USB Connector Unpacking and Inspection of the Instrument Remove all items from packing carton and carefully inspect to ensure that no visible damage has occurred during shipment and that all items listed below have been received If the items received do not match the order please ...

Страница 5: ...igned to provide the ability to customize the instrument according to your needs at any time during normal operation The Setup screen has been split into sub menus to facilitate instrument configuration This section describes how to use each of the sub menus to configure the instrument Turning on the Instrument 1 Confirm that the M100 is safely and properly plugged into a suitable power source 2 T...

Страница 6: ...ration The screen has been factory adjusted but can be adjusted if the alignment is off The screen may need to be calibrated if new software has been downloaded It is a 5 position setup 1 Press the Setup icon Select Touch Screen Calibration 2 If this calibration fails restart the instrument by unplugging and reinserting the power plug 3 Press the top screen time banner for 5 seconds to reset to fa...

Страница 7: ...nsert the USB memory stick that contains the software into the USB port 3 If the firmware revision on the memory stick is older than the firmware currently installed the instrument will ask if you wish to continue 4 Press the to continue with the firmware installation The upload will take several minutes During the initial upload the screen will show busy and screen anomalies may appear until the ...

Страница 8: ...ided as general guidelines for proper care of ProCal standards Standards should only be handled by the top black cap or by the very bottom of the cuvette Keep cuvette glass clean of fingerprints and debris Cuvettes can be cleaned with any domestic glass cleaner and the provided microfiber cloth Prevent ProCal standards from freezing This will most likely destroy them Standards under 100 NTU must b...

Страница 9: ...on on the left side of the screen 2 Insert the 4000 NTU standard 3 Press the Command button in the middle of the screen to start calibration A countdown from 30 will appear on the screen When the countdown is complete the screen will request the 1000 NTU standard 4 Press the 1000 NTU button on the left side of the screen 5 Insert the 1000 NTU standard 6 Press the Command button to continue the cal...

Страница 10: ...ostic Banner On Off Reading List Setup Calibration Trend Home Data Retrieval Data logging can only save 100 points per Sample ID All data is saved in CSV format and is date and time stamped Sample ID Data can be downloaded to a USB memory stick or printed in the Trend screen after pressing Download Print Calibration data can be downloaded to a USB memory stick in the Calibration Download menu by p...

Страница 11: ...Toll Free 888 203 7248 Fax 239 454 0694 Email HF Info wattswater com www hfscientific com Routine Maintenance Routine maintenance should include calibration of the instrument See Calibration section in this manual for more information Cleaning the Cuvette Measurement cuvettes used for the sample should be clean and free of marks or scratches Clean the interior and exterior of the cuvettes with a d...

Страница 12: ...e cost of repairing or replacing other property which is damaged if this product does not work properly other costs resulting from labor charges delays vandalism negligence fouling caused by foreign material damage from adverse water conditions chemical or any other circumstances over which the Company has no control In addition the Company shall not be responsible for any costs incidental to the ...

Отзывы: